Sterilization parameters: times, temperatures and monitoring
The headline steam numbers from CDC are 121°C (250°F) for at least 30 minutes in a gravity-displacement sterilizer, or 132°C (270°F) for at least 4 minutes in a prevacuum sterilizer, for wrapped supplies. Every other method has its own time, temperature and spore test, and the device maker's instructions can call for more.
Updated November 8, 2026
These are the minimums published in federal guidance. Your sterilizer's and your device's instructions for use (IFU) come first: when they ask for a longer cycle, run the longer cycle.
Steam sterilization: minimum cycle times for wrapped items
| Cycle | Temperature | Minimum exposure | Notes |
|---|---|---|---|
| Gravity displacement, wrapped instruments | 121°C (250°F) | 30 min | Drying time 15–30 min |
| Gravity displacement, wrapped instruments | 132°C (270°F) | 15 min | Drying time 15–30 min |
| Gravity displacement, textile packs | 121°C (250°F) | 30 min | Drying time 15 min |
| Gravity displacement, textile packs | 132°C (270°F) | 25 min | Drying time 15 min |
| Gravity displacement, wrapped utensils | 121°C (250°F) | 30 min | Drying time 15–30 min |
| Gravity displacement, wrapped utensils | 132°C (270°F) | 15 min | Drying time 15–30 min |
| Prevacuum (dynamic air removal), wrapped instruments | 132°C (270°F) | 4 min | Drying time 20–30 min. No 121°C time is listed for prevacuum. |
| Prevacuum (dynamic air removal), textile packs | 132°C (270°F) | 4 min | Drying time 5–20 min |
| Prevacuum (dynamic air removal), wrapped utensils | 132°C (270°F) | 4 min | Drying time 20 min |
| Steam-flush pressure-pulse, porous loads and instruments | 132–135°C | 3–4 min | Typical values. Air is pushed out with steam above atmospheric pressure, so air leaks do not affect it. |
- CDC adapted Table 7 from AAMI. These are minimums; a device, container or sterilizer IFU can require a longer exposure. Follow the manufacturer's instructions.
- Steam should be dry saturated steam with a dryness fraction of at least 97%.
- Gravity sterilizers leave some air in the load, which slows steam getting into porous items. CDC notes 10 lb of microbiological waste needs at least 45 minutes at 121°C.
Immediate-use steam sterilization (IUSS, called 'flash' in the CDC guideline)
| Cycle | Temperature | Minimum exposure | Notes |
|---|---|---|---|
| Gravity displacement | 132°C (270°F) | 3 min | Nonporous items only: routine metal instruments, no lumens |
| Gravity displacement | 132°C (270°F) | 10 min | Nonporous and porous items together, such as rubber, plastic or lumened items |
| Prevacuum (dynamic air removal) | 132°C (270°F) | 3 min | Nonporous items only: routine metal instruments, no lumens |
| Prevacuum (dynamic air removal) | 132°C (270°F) | 4 min | Nonporous and porous items together, such as rubber, plastic or lumened items |
| Steam-flush pressure-pulse | 132°C (270°F) | 4 min | Nonporous or mixed loads. Follow the manufacturer's instructions. |
- CDC labels Table 8 as examples adapted from AAMI. The sterilizer and device IFUs set the actual cycle.
- Flash was first defined as an unwrapped item at 132°C and 27–28 lb of pressure, held 3 minutes, in a gravity sterilizer.
- CDC: do not flash sterilize implants unless it cannot be avoided, and never use it for convenience or to save time.
Dry heat sterilization
| Cycle | Temperature | Minimum exposure | Notes |
|---|---|---|---|
| Hot air sterilizer | 170°C (340°F) | 60 min | Shortest common setting |
| Hot air sterilizer | 160°C (320°F) | 120 min | |
| Hot air sterilizer | 150°C (300°F) | 150 min | Longest common setting |
- Monitor with Bacillus atrophaeus spores, which resist dry heat better than Geobacillus stearothermophilus.
- Dry heat kills slowly and the heat damages most materials, so it suits only items steam cannot reach or would harm.
Source: CDC, Disinfection and Sterilization Guideline: Other Sterilization Methods
Ethylene oxide (EtO) sterilization
| Parameter | Range or value | Notes |
|---|---|---|
| Temperature | 37–63°C | Operating range |
| Gas concentration | 450–1200 mg/L | Operating range |
| Relative humidity | 40–80% | Water molecules carry the gas to its targets |
| Exposure time | 1–6 hours | Operating range |
| Whole cycle | About 2.5 hours | Not counting aeration. Five stages: preconditioning and humidification, gas in, exposure, evacuation, air washes. |
| Mechanical aeration | 8–12 hours at 50–60°C | Lets absorbed EtO leave the materials |
| Room-air aeration | 7 days at 20°C | Works but is slow |
| Biological indicator | Bacillus atrophaeus | |
| OSHA exposure limits | 1 ppm (8-hour TWA); action level 0.5 ppm; excursion limit 5 ppm over 15 min | Worker safety limits, not cycle settings |
- These are the operating ranges CDC gives. The settings for a given load come from the sterilizer and device IFUs.
- Many materials absorb EtO, so items must be aerated before use. Follow the device maker's aeration instructions.
Source: CDC, Disinfection and Sterilization Guideline: Ethylene Oxide Sterilization
Hydrogen peroxide gas plasma
| Parameter | Range or value | Notes |
|---|---|---|
| Temperature | 37–44°C | Low-temperature process |
| Peroxide in the chamber | 6 mg/L | Vaporized from a cassette, as described by the maker |
| Cycle time, original system | 75 min | |
| Cycle time, two-stage version | 52 min | Software change cut it from 73 min |
| Cycle time, newest version | 28–38 min | Check the maker's device size and lumen limits |
| Aeration | None needed | By-products are water vapor and oxygen |
| Moisture | Load must be dry | Wet items stop the vacuum and the cycle aborts |
| Packaging | No linen or paper | Use packaging made for the process |
| Biological indicator | Geobacillus stearothermophilus | Per CDC's monitoring section; see note |
- CDC's monitoring section lists G. stearothermophilus for hydrogen peroxide gas plasma, but the gas plasma chapter says one system used B. atrophaeus spores. Use the BI the sterilizer maker specifies.
- Cycle times are from older and newer models of one system. Follow the manufacturer's instructions for your unit.
Vaporized hydrogen peroxide
| Parameter | Range or value | Notes |
|---|---|---|
| Liquid peroxide strength | 30–35% | Drawn from a disposable cartridge |
| Cycle time | 30–45 min | Example of a rapid cycle |
| Biological indicator | Not named by CDC | Follow the manufacturer's instructions |
- CDC gives few numbers for this method. Load limits, lumen limits and cycle choice come from the sterilizer IFU.
Source: CDC, Disinfection and Sterilization Guideline: Other Sterilization Methods
Liquid peracetic acid (immersion system)
| Parameter | Range or value | Notes |
|---|---|---|
| Concentrate | 35% peracetic acid | Single-dose container, opened right before the cycle |
| Use strength | 0.2% | Diluted with filtered water |
| Temperature | About 50°C | |
| Exposure | 12 min | Sterilant circulates in the chamber and through scope channels |
| After the cycle | Use at once | Point-of-use system with no sterile storage |
| Biological indicator | Geobacillus stearothermophilus |
- Flexible scopes need the right channel connectors so sterilant reaches every channel.
- The spent solution goes down the drain after each cycle.
Biological indicator organism by method
| Method | BI organism | Incubation | Notes |
|---|---|---|---|
| Steam | Geobacillus stearothermophilus (10^5 spores) | 55–60°C | |
| Ethylene oxide | Bacillus atrophaeus (10^6 spores) | 35–37°C | |
| Dry heat | Bacillus atrophaeus (10^6 spores) | 35–37°C | |
| Hydrogen peroxide gas plasma | Geobacillus stearothermophilus (10^5 spores) | 55–60°C | Use the BI the sterilizer maker specifies |
| Liquid peracetic acid | Geobacillus stearothermophilus (10^5 spores) | 55–60°C |
- Run BIs at least weekly on steam and low-temperature sterilizers and in every load with an implant. CDC notes daily BIs on a unit that runs several loads a day catch problems sooner.
- Rapid-readout steam BIs read in 1 hour for 132°C gravity cycles and 3 hours for 121°C gravity and 132°C vacuum cycles.
Source: CDC, Disinfection and Sterilization Guideline: Sterilizing Practices
How each load is monitored
- Check every load with mechanical monitors (time, temperature, pressure) and with chemical indicators inside and outside the pack; if the inside indicator can be seen, no outside one is needed. source
- Physical monitoring of a steam cycle means reading the printout or graph of temperature, time at temperature and pressure. source
- Each day, check a steam sterilizer's cycle time and temperature on its record chart or printout, and its pressure on the gauge. source
- An outside chemical indicator only shows the pack went through a cycle; an inside indicator shows the sterilant got in. Neither proves sterility. source
- CDC says chemical indicators fall into five classes by how many sterilization variables they respond to, but it does not name the classes. source
- FDA's 2003 guidance for chemical indicator makers names three types: process indicators for individual packs, chemical integrators that respond to all critical variables, and air-removal indicators used in a Bowie-Dick test pack. source
- Run the Bowie-Dick (air-removal) test daily on any prevacuum sterilizer in use, ahead of the first load, in an empty chamber at 134°C for 3.5 minutes. source
- Lay the Bowie-Dick pack flat in the front, lower part of the rack, by the door and above the drain. An even color change passes; a spot means trapped air, and the sterilizer stays out of use until it is inspected and passes. source
- Test steam and low-temperature sterilizers with BIs at least weekly. CDC adds that daily BIs on a unit running several loads a day catch failures sooner. source
- Put a BI in every load that holds an implant, and hold the implants until the BI reads negative whenever possible. source
- For routine steam BI testing, place the test pack flat in a chamber that is otherwise full, at the hardest spot, usually front bottom near the drain, and incubate an unexposed control BI from the same lot. source
- A standard full-cycle steam BI does not reliably monitor flash cycles; use BIs made for flash sterilization. source
- After installation, relocation, redesign or major repair, run three empty steam cycles in a row with a BI and CI test pack, plus three Bowie-Dick cycles in a row for a prevacuum unit, before putting it back in use. source
- Release a load only if the mechanical and chemical results are acceptable; do not use items when either suggests poor processing. source
- Steam, one positive BI: non-implant items need not be recalled unless maintenance finds the sterilizer or procedure at fault; retest at once, and if BIs stay positive, recall and reprocess the implicated loads. source
- Other methods (such as EtO or gas plasma), one positive BI: treat everything run since the last negative BI, up to the next good BI, as nonsterile and retrieve it for reprocessing if possible. source
- Label each sterilized item with a load number showing the sterilizer, the cycle or load number and the date, plus an expiration date if one is used. source
- Keep mechanical, chemical and biological sterilization records as long as standards and state and federal rules require. CDC gives the Joint Commission's 3-year request as an example. source
Flashcards: 34 cards on these numbers
Take the free CRCST practice test
Common questions
What are the standard steam sterilization parameters?
For wrapped supplies, CDC lists at least 30 minutes at 121°C (250°F) in a gravity-displacement sterilizer, or at least 4 minutes at 132°C (270°F) in a prevacuum sterilizer. These are minimums, and the device or sterilizer IFU can call for longer.
What are gravity sterilization parameters?
In CDC's table, every wrapped item in a gravity cycle at 121°C needs 30 minutes. At 132°C, wrapped instruments and utensils need 15 minutes and textile packs need 25 minutes, with 15 to 30 minutes of drying.
What are flash (IUSS) sterilization parameters?
CDC's examples are all at 132°C (270°F). Metal items with no lumens need 3 minutes in either gravity or prevacuum cycles; mixed loads with porous or lumened items need 10 minutes by gravity or 4 minutes by prevacuum.
What are EtO sterilization parameters?
CDC gives operating ranges of 37–63°C, 450–1200 mg/L of gas, 40–80% relative humidity and 1–6 hours of exposure. The cycle takes about 2.5 hours, then items need 8–12 hours of mechanical aeration at 50–60°C.
What are plasma sterilization parameters?
CDC describes hydrogen peroxide gas plasma running at 37–44°C, with cycle times from 75 minutes on the first model down to 28–38 minutes on the newest. No aeration is needed, but the load must be dry and free of linen or paper.
What biological indicator organism is used for each sterilization method?
CDC lists Geobacillus stearothermophilus (10^5 spores) for steam, liquid peracetic acid and hydrogen peroxide gas plasma. Bacillus atrophaeus (10^6 spores) is used for ethylene oxide and dry heat.
What are the sterilization indicator classes?
CDC sorts chemical indicators into five classes by how many sterilization variables they detect, but it does not name them. An outside indicator shows the pack was processed and an inside one shows the sterilant got in; a BI is still needed because only spores measure killing power.
How often do you run a Bowie-Dick test?
Daily on any prevacuum steam sterilizer in use, ahead of the first load, in an empty chamber. CDC describes running it at 134°C for 3.5 minutes; if it fails, the sterilizer is not used until it is inspected and passes.