Study these as flashcardsSterilization parameters
Work Environment, Air and Chemical Safety
- Negative pressure:
- A room held at lower air pressure than its neighbors, so air flows into it; used for decontamination.
- Positive pressure:
- A room held at higher air pressure than its neighbors, so air flows out of it; used for sterile storage.
- Air changes per hour:
- How many times the full volume of a room's air is replaced in one hour.
- Relative humidity:
- The amount of moisture in the air, given as a percentage of the most the air could hold at that temperature.
- Smoke tube:
- A device that releases a visible wisp at a doorway so staff can see which way air is moving.
- Flutter strip:
- A light paper or plastic strip hung at a door gap that bends in the direction of airflow.
- Safety Data Sheet (SDS):
- A 16-section document from the maker of a chemical describing its hazards, safe handling, first aid, spill response and disposal.
- Hazard Communication Standard:
- OSHA rule 29 CFR 1910.1200 requiring chemical labels, data sheets and worker training about chemical hazards.
- Workplace container:
- A bottle or other vessel a worker fills from the original product for use on the job; it must be labeled.
- Product identifier:
- The name or number on a chemical's label that links it to its matching data sheet.
- Eyewash station:
- Plumbed or self-contained equipment that flushes the eyes with a steady stream of water after a chemical splash.
- Glutaraldehyde:
- A high-level disinfecting chemical that can sting the eyes, inflame the skin and irritate the airways.
Attire, Traffic Control and Ergonomics
- Hand hygiene:
- Cleaning the hands by washing with soap and water or rubbing in an alcohol-based product.
- Alcohol-based hand sanitizer:
- A rub-in hand product preferred by CDC when hands are not visibly soiled; rub until dry, about 20 seconds.
- Visibly soiled hands:
- Hands with dirt, blood or other matter you can see; they must be washed with soap and water.
- Gram-negative bacteria:
- A large group of bacteria, including Pseudomonas and Klebsiella, found more often on the fingertips of workers wearing artificial nails.
- Bloodborne Pathogens standard:
- OSHA rule 29 CFR 1910.1030 that protects workers exposed to blood and other potentially infectious materials.
- Traffic control:
- Limiting who enters each area and how people and items move, so contamination does not travel from dirty to clean.
- One-way workflow:
- Moving items only from decontamination to packaging to sterilization to storage, never back toward dirty areas.
- Limited-access area:
- A space, such as sterile storage, entered only by staff who need to work there.
- Ergonomics:
- Fitting a job to a person to reduce fatigue and work-related muscle, nerve and tendon injuries.
- Musculoskeletal disorder (MSD):
- An injury or illness of the muscles, nerves, tendons, ligaments or blood vessels, often from repeated strain.
- Engineering control:
- A physical change to the workplace, such as an adjustable table, that removes or reduces a hazard; OSHA's top choice.
- Administrative control:
- A rule or procedure, such as job rotation or scheduled breaks, used when physical changes are not enough.
- Power zone:
- The safest lifting area: between knee and shoulder height, with the load held near the body.
Regulations, Standards, Records and Quality
- FDA:
- Federal agency that regulates medical device makers and clears sterilants and high-level disinfectants for devices.
- EPA:
- Federal agency that registers disinfectants for environmental surfaces such as floors and countertops.
- OSHA:
- Federal agency that sets and enforces rules to keep workers safe and healthy on the job.
- Conditions of Participation:
- CMS health and safety requirements an organization must meet to take part in Medicare and Medicaid.
- Deemed status:
- Recognition that a facility accredited by a CMS-approved organization meets Medicare requirements.
- Consensus standard:
- A voluntary practice document written by a standards development organization such as AAMI.
- Instructions for use (IFU):
- The manufacturer's validated written directions for using, cleaning and processing a specific device.
- FIFRA:
- The federal pesticide law under which EPA registers surface disinfectants and makes their labels legally binding.
- Load number:
- A code on each package linking it to the sterilizer cycle that processed it, for tracing and recalls.
- MedWatch:
- FDA's program for safety information and adverse event reports; voluntary reports use Form FDA 3500.
- User facility reporting:
- FDA rule requiring hospitals to report device-related deaths and serious injuries within 10 work days.
- Competency testing:
- Documented checks that a worker can do each task correctly, done at hire and regularly after.
- PDSA:
- Plan-Do-Study-Act: an improvement cycle built on small tests of change before wider adoption.
- Lean:
- An improvement method that raises efficiency by removing steps that add no value.
- Six Sigma:
- A data-driven improvement method that reduces variation and defects using Define-Measure-Analyze-Improve-Control.
- Root cause analysis (RCA):
- A structured review of a serious event to find underlying system problems instead of individual blame.
Decontamination Safety, PPE and Microbiology
- Chain of infection:
- Five linked steps (reservoir, exit, transmission, entry, host) that an infection needs; breaking one stops it.
- Fomite:
- Inanimate object, such as a used instrument, that carries organisms from one person to another.
- Indirect transmission:
- Spread of an infectious agent through a go-between such as an object, air particle or insect.
- Biofilm:
- Community of microbes tightly attached to a surface, far more resistant to germicides than free-floating cells.
- Bacterial spore:
- Dormant bacterial form protected by a coat and cortex; the most chemical-resistant organism after prions.
- Mycobacteria:
- Bacteria such as M. tuberculosis whose waxy cell wall makes them resist many disinfectants.
- Exposure incident:
- Blood or infectious material reaching eye, mouth, mucous membrane, broken skin or a puncture during work.
- Engineering control:
- Device that isolates or removes a hazard, such as a sharps container.
- Work practice control:
- Change in how a task is done to cut exposure, such as no eating in the work area.
- Hepatitis B declination:
- Signed form refusing the HBV vaccine; the worker may still accept the shots later, free.
- Post-exposure evaluation:
- Confidential, free medical assessment and follow-up the employer provides after an exposure incident.
- Biohazard label:
- Fluorescent orange or orange-red warning tag with the symbol; red bags or containers may substitute.
- Donning:
- Putting on protective gear in order: gown, mask, eye protection, then gloves over the cuffs.
- Doffing:
- Removing protective gear; in CDC's first example gloves come off first and the mask last.
- Prion:
- Infectious protein that causes CJD and is not reliably inactivated by routine reprocessing.
- High-infectivity tissue:
- For CJD: brain, spinal cord and eye, which call for special instrument reprocessing.
Manual and Mechanical Cleaning
- Point-of-use treatment:
- Keeping used instruments open and moist right after the procedure; it starts but never replaces cleaning.
- Friction:
- Rubbing or scrubbing soil off with a brush or cloth; one of two essentials of manual cleaning.
- Fluidics:
- Fluid forced under pressure through channels to flush out soil, especially where a brush cannot pass.
- Enzymatic cleaner:
- Detergent with proteases (and sometimes lipases or amylases) that digests soil but kills no microbes.
- Neutral pH detergent:
- Cleaning solution that is neither acidic nor alkaline; with enzymes it is CDC's preferred choice for delicate instruments and flexible endoscopes.
- Alkaline cleaner:
- High-pH detergent that dissolves protein and fat efficiently but can corrode devices.
- Cavitation:
- Formation and implosion of tiny bubbles from sound waves, which pulls debris off instruments in an ultrasonic unit.
- Degassing:
- Running a freshly filled ultrasonic unit with the lid closed for 5 to 10 minutes to remove dissolved gas.
- Impingement:
- Pressurized water spray that physically knocks soil off; the main cleaning action of a washer-disinfector.
- Thermal rinse:
- Final hot-water step of a washer-disinfector cycle that disinfects by heat.
- Test soil:
- Commercial product mimicking dried blood, used to check that a washer-disinfector is cleaning properly.
- Cart washer:
- Machine for carts, rigid containers and basins; lacks enzymatic wash and lubrication steps.
- Pyrogen-free water:
- Water free of fever-causing substances such as endotoxin; required for final rinsing of intravascular device lumens.
- TASS:
- Toxic anterior segment syndrome: acute, noninfectious eye inflammation after anterior segment surgery, often from instrument residue.
- Single-use device:
- Item intended for one use or one patient in one procedure, then discarded.
- Instructions for use (IFU):
- Manufacturer's written directions for reprocessing a device or using a chemical; they set disassembly, dilution, temperature and method.
Spaulding Classification and Disinfection
- Critical item:
- Device entering sterile tissue or the vascular system, such as a surgical instrument; must be sterile.
- Semicritical item:
- Device touching mucous membranes or nonintact skin, such as a laryngoscope blade; needs at least high-level disinfection.
- Noncritical item:
- Device touching only intact skin, such as a blood pressure cuff; needs low-level disinfection.
- High-level disinfection:
- Process killing all microorganisms except small numbers of bacterial spores.
- Intermediate-level disinfectant:
- Chemical that kills vegetative bacteria including tubercle bacilli, fungi, lipid and some nonlipid viruses, but not spores.
- Low-level disinfectant:
- Chemical that kills most vegetative bacteria, lipid viruses and some fungi, but not tubercle bacilli or spores.
- Chemical sterilant:
- Germicide that kills spores when given long contact, often 3 to 12 hours.
- Minimum effective concentration (MEC):
- Lowest strength at which a reusable disinfectant still meets its label claim; checked with test strips.
- Reuse life:
- Maximum number of days a reusable solution may be used, set by its manufacturer, regardless of strip results.
- Glutaraldehyde:
- Metal-safe high-level disinfectant whose vapor irritates airways and skin; needs ventilation and nitrile or butyl gloves.
- Ortho-phthalaldehyde (OPA):
- High-level disinfectant that stains proteins, including unprotected skin, gray.
- Peracetic acid:
- Sporicidal chemical leaving no residue that can corrode copper, brass, bronze and plain steel.
- Sodium hypochlorite:
- Bleach; fast and cheap but corrosive at high strength and releases chlorine gas with ammonia or acid.
- Quaternary ammonium compound:
- Low-level surface disinfectant for noncritical items; not sporicidal and generally not tuberculocidal.
- FIFRA:
- Federal law under which EPA registers disinfectants; using a product against its label is a violation.
- Phenolic:
- EPA-registered surface disinfectant, not cleared for high-level use, kept off occupied infant bassinets.
Instrument Inspection, Testing and Care
- Box lock:
- The hinged joint where the two halves of a clamp or hemostat cross; a common place for hidden soil and cracks.
- Serrations:
- Grooves or ridges cut into instrument jaws to improve grip; they trap debris and need close inspection.
- Ratchet:
- Interlocking teeth near the finger rings that hold an instrument's jaws closed at set positions.
- Function check:
- Testing that an instrument works as intended, such as jaw alignment, ratchet hold, smooth hinges and sharp edges.
- Retained soil:
- Blood, tissue or other debris left on a device after cleaning, which can shield microbes from sterilization.
- Insulation tester:
- A device that detects breaks in the coating of electrosurgical instruments that could let current escape.
- Cannulated device:
- An instrument with a hollow channel, such as a suction tip, checked to confirm the channel is open.
- Passive layer:
- The thin chromium and iron oxide film on stainless steel that protects it from rust.
- Pitting:
- Small dark holes where metal has been lost to corrosion, often from blood, chlorides or harsh solutions.
- Spotting:
- Loose deposits sitting on an instrument's surface, often minerals from hard water, that can be wiped away.
- Galvanic corrosion:
- Metal transferring from one metal to another when dissimilar metals share a solution, leaving rainbow, gold or rusty tints.
- Water-soluble lubricant:
- An instrument lubricant that dissolves in water, so it does not block sterilant from reaching the metal.
- Instructions for use (IFU):
- The manufacturer's written directions for cleaning, testing, lubricating and sterilizing a specific device.
- Deionized water:
- Water with dissolved minerals removed, used for final rinsing to prevent spots and deposits.
Set Assembly and Chemical Indicators
- Tray list:
- The count sheet naming every device a set should contain, used to check completeness during assembly.
- Stringer:
- A pin or rack that holds ringed instruments together, open, inside a tray.
- First ratchet:
- The lightest locked position on a ratcheted instrument, the only closed position allowed in a set.
- Tip protector:
- A cover over a sharp end, validated for steam penetration and fitted loosely enough for sterilant to reach.
- Wet pack:
- A package with moisture inside the case or tray after the cycle ends; its contents are considered contaminated.
- External chemical indicator:
- A marker on the outside of a package showing it went through a cycle, not that it is sterile.
- Internal chemical indicator:
- A marker placed inside a package to show the sterilant reached the contents.
- Process indicator:
- ISO 11140 Type 1; shows a pack was exposed to the process, separating processed from unprocessed units.
- Integrating indicator:
- ISO 11140 Type 5; reacts to all critical variables and is meant to equal or exceed biological indicator performance requirements.
- Emulating indicator:
- ISO 11140 Type 6; reacts to all critical variables and matches those of a specified cycle.
- Bowie-Dick test:
- A daily empty-chamber check of air removal and air leaks in a prevacuum steam sterilizer.
- Biological indicator:
- A test containing resistant spores; the only monitor that measures whether a cycle killed microbes.
- Quarantine:
- Holding processed items back from use until a pending test result, such as a spore test, comes back.
- Loaner set:
- Instruments borrowed from a vendor or another facility for a procedure, processed here per the maker's instructions before use.
- Immediate-use steam sterilization:
- A shortened steam cycle for items needed at once; the WHO manual says never to use it for implants.
Packaging Methods and Labeling
- Sterile barrier:
- Packaging that lets sterilant in, then blocks microbes so contents stay sterile until opened.
- Sequential wrapping:
- Two sheets wrapped one after the other, creating a package within a package.
- Simultaneous wrapping:
- Two sheets laid together and folded as one, giving two layers with a single wrapping step.
- Envelope fold:
- A wrapping method with the set placed diagonally, slightly off center, and the corners folded over it.
- Square fold:
- A wrapping method, also called parcel fold, with the set centered and parallel to the wrap's long edge.
- Tenting:
- Loose wrap standing away from the contents, leaving air pockets that can compromise the barrier.
- Peel pouch:
- A heat-sealed paper-plastic or synthetic package for a single, lightweight device.
- Heat sealer:
- A machine that closes pouches with heat and pressure; tested daily and calibrated at least yearly.
- Rigid container:
- A reusable metal or polymer box with filtered vents, used mainly for large steam-sterilized sets.
- Tyvek:
- A synthetic polyolefin material with low chemical absorption, suitable for hydrogen peroxide plasma packaging.
- Load number:
- Label code identifying the sterilizer, the cycle or load, the date and any expiration date.
- Event-related shelf life:
- Treating an item as sterile until damage, moisture, a broken seal or other event compromises it.
- Batch control label:
- A peel-off label recording lot details, placed in the patient's record to allow tracing and recall.
- Indicator tape:
- Adhesive tape that closes wrapped packs and carries a process indicator that changes color in a cycle.
Steam and High-Temperature Sterilization
- Gravity displacement:
- Air removal in which incoming steam, lighter than air, pushes air down and out through a drain near the bottom.
- Dynamic air removal:
- Active air removal before exposure, either by vacuum pulses (prevacuum) or by steam flushes and pressure pulses.
- Prevacuum sterilizer:
- Steam unit that uses a vacuum pump or ejector to pull air out of the chamber and load before exposure.
- Steam-flush pressure-pulse:
- Air-removal method using repeated steam flushes and pulses above atmospheric pressure, so leaks cannot draw air in.
- Conditioning:
- First cycle phase, when air is removed and the load is heated.
- Exposure time:
- Period the load is held at the set sterilizing temperature.
- Jacket:
- Steam-filled space around the chamber that preheats it and helps dry the load.
- Door gasket:
- Seal between the door and chamber that keeps steam and pressure inside.
- Thermostatic trap:
- Drain-line device that lets condensate leave the chamber, helping loads dry.
- Drain strainer:
- Screen over the chamber drain that catches lint and debris; clean it daily.
- Dryness fraction:
- Share of steam that is vapor rather than liquid water; at least 97% is the goal.
- Wet pack:
- Package with dampness, droplets or puddles on or inside it after the cycle; treated as contaminated.
- Immediate-use steam sterilization:
- Unwrapped steam processing of an item needed right away; also called flash; no storage afterward.
- Porous item:
- Item that can trap air or liquid, such as fabric, rubber, plastic or a device with a lumen.
Low-Temperature Sterilization
- Ethylene oxide (EtO):
- Flammable, carcinogenic sterilant gas used for heat- and moisture-sensitive devices; requires aeration afterward.
- Aeration:
- Holding items after an EtO cycle, in an aerator or room, until the gas they absorbed has left them.
- Preconditioning:
- First EtO cycle stage, when the load is brought to set conditions and humidified before gas enters.
- Permissible exposure limit (PEL):
- OSHA's maximum allowed worker exposure; for EtO, 1 ppm averaged over 8 hours.
- Excursion limit:
- OSHA's short-term EtO cap: 5 ppm averaged over 15 minutes.
- Action level:
- EtO exposure of 0.5 ppm over 8 hours; reaching it 30 or more days a year requires medical surveillance.
- Time-weighted average (TWA):
- A worker's average airborne chemical concentration across a set sampling period, such as an 8-hour shift.
- Hydrogen peroxide gas plasma:
- Low-temperature vacuum process using peroxide vapor and plasma; leaves water and oxygen, so no aeration.
- Cellulose:
- Plant-based material such as paper or cotton linen; absorbs hydrogen peroxide, so it cannot go in peroxide sterilizers.
- Polyolefin pouch:
- Synthetic (Tyvek-type) peel pouch acceptable for hydrogen peroxide sterilization.
- Peracetic acid system:
- Automated liquid immersion sterilizer for immersible heat-sensitive instruments; items must be used immediately.
- Channel connector:
- Fitting that links an endoscope's lumens to the processor so liquid sterilant flows through them.
- Ozone sterilizer:
- Low-temperature unit that generates its oxidizing gas from oxygen, water and electricity, then converts it back with a catalyst.
- Dry heat:
- Hot-air sterilization, e.g. 170°C for 60 minutes, for powders, oils and petroleum products.
Sterilizer Testing, Load Release and Recalls
- Physical monitor:
- Printout, graph or display showing a cycle's time, temperature and pressure; also called mechanical monitoring.
- External chemical indicator:
- Indicator on a package's outside showing only that it was exposed to a sterilization process.
- Internal chemical indicator:
- Indicator placed inside a package to show sterilant reached the contents.
- Process indicator:
- ISO Type 1 marker, such as tape, separating processed from unprocessed packages.
- Integrating indicator:
- ISO Type 5 device reacting to all critical variables, designed to perform at or beyond BI requirements.
- Emulating indicator:
- ISO Type 6 device reacting to all critical variables, matched to the specific cycle on its label.
- Bowie-Dick test:
- Daily empty-chamber air-removal check for prevacuum steam sterilizers, run before the first load.
- Leak test:
- Vacuum check of whether a prevacuum chamber is airtight, run on the sterilizer maker's schedule.
- Biological indicator (BI):
- Carrier of resistant spores that shows whether a cycle actually killed organisms.
- Geobacillus stearothermophilus:
- Spore-former used to monitor steam (and per CDC peroxide plasma and peracetic acid); incubated at 55-60°C.
- Bacillus atrophaeus:
- Spore-former used to monitor EtO and dry heat; incubated at 35-37°C.
- Control BI:
- Unprocessed indicator from the same lot, incubated to prove the spores could grow.
- Process challenge device (PCD):
- Test pack or tray that challenges the cycle and holds a BI, integrating indicator or enzyme-only indicator.
- Quarantine:
- Holding processed items, such as implants, until monitoring results allow release.
- Recall:
- Retrieving items from suspect loads so they can be reprocessed and patient exposure assessed.
- Load label:
- Package marking naming the sterilizer, cycle or load number, date and any expiration, enabling tracing.
Sterile Storage, Shelf Life and Transport
- Sterile storage:
- A limited-access area with controlled temperature and humidity where processed packs are kept until issue.
- Event-related shelf life:
- The practice of treating a pack as sterile until damage, wetting or another event compromises it.
- Time-related shelf life:
- The practice of labeling packs with an expiration date and reprocessing them once that date passes.
- Package integrity:
- The condition of a wrap, pouch or container being whole, dry and sealed, with no tears or punctures.
- Impervious packaging:
- Material, such as heat-sealed plastic, that does not let moisture or microbes pass through.
- Stock rotation:
- Placing newer items behind older ones so the oldest are used first; also called first in, first out.
- First in, first out (FIFO):
- Using the oldest stock of an item before newer deliveries of the same item.
- Open shelving:
- Uncovered racks for supplies; allowed by CDC, though closed or covered cabinets are ideal.
- Sprinkler clearance:
- The 18-inch space CDC calls for between stored items and a ceiling fire sprinkler head.
- Corrugated cardboard:
- Ridged material used for shipping boxes; such boxes are never taken into sterile storage areas.
- Closed cart:
- A transport cart with walls, doors or a cover that protects its load while outside the department.
- Reprocessing:
- Repeating cleaning, inspection, packaging and sterilization so an item can be used safely again.
Inventory, Supply Chain and Distribution
- PAR level:
- The set quantity of an item a location should hold; restocking brings it back to that number.
- Exchange cart:
- One of a set of identical stocked carts; a full one is swapped in while the used one is refilled.
- Case cart:
- A cart holding the instruments and supplies for one surgical procedure, which then returns the used items.
- Code cart:
- A crash cart of emergency drugs and equipment, stocked to the facility's approved contents list.
- Just-in-time inventory:
- Receiving supplies only as they are needed, which lowers stock on hand but leaves little backup.
- Consignment inventory:
- Supplier-owned stock kept at the facility, paid for only when an item is used.
- Inventory turnover:
- How quickly stock is used and replaced; slow movement points to overstock and expiring items.
- Unit of measure:
- The packaging quantity an item is counted and issued in, such as each, box or case.
- Unique Device Identifier (UDI):
- An FDA-required code on device labels that lets a device be identified from distribution through use.
- Device identifier (DI):
- The fixed part of a device's label code naming the labeler and the model or version.
- Production identifier (PI):
- The variable part of a device's label code: lot, serial number, expiration and manufacture dates.
- GUDID:
- FDA's Global Unique Device Identification Database, which stores the device-identifier portion of each UDI.
- AIDC:
- Automatic identification and data capture: machine-readable formats such as barcodes.
- Class I recall:
- FDA's most serious recall category: a reasonable chance of serious health problems or death.
- Correction:
- A recall action that fixes a device problem where the device is used, without removing it.
- Quarantine:
- Holding processed items back from use until test results, such as a biological indicator, come back negative.
Patient Care Equipment
- Noncritical item:
- Equipment that touches only intact skin or no patient at all; needs cleaning plus low-level disinfection.
- Spaulding classification:
- A system sorting devices as critical, semicritical or noncritical by infection risk, which sets the processing level.
- Low-level disinfection:
- Killing most vegetative bacteria and some fungi and viruses; the level CDC recommends for noncritical items.
- Contact time:
- How long a disinfectant must stay wet on a surface to work, as stated on its label.
- EPA-registered disinfectant:
- A surface germicide whose label claims and directions are approved by the Environmental Protection Agency and legally binding.
- Cross-contamination:
- Spread of microbes from soiled items or surfaces to clean ones.
- Single-patient-use item:
- A part, such as a compression sleeve, meant for one patient and discarded rather than reprocessed.
- Biomedical (Clinical) Engineering:
- The hospital department that inspects, maintains and repairs medical equipment.
- Preventive maintenance (PM):
- Scheduled inspection, testing and servicing of equipment by qualified staff, documented on a label or record.
- Grounding prong:
- The third plug pin that keeps equipment connected to ground; altering it or bypassing it with an adapter is prohibited.
- Function check:
- A test before issue that equipment powers on and works as its manufacturer's instructions describe.
- Equipment inventory:
- A list of all hospital equipment, owned or leased, with each item's maintenance record.
Communication, Teamwork and Professional Growth
- Protected health information (PHI):
- Identifiable information about a person's health, care or payment for care, in any form: paper, electronic or spoken.
- HIPAA:
- The federal law whose Privacy Rule sets how health care organizations must protect identifiable patient information.
- Minimum necessary:
- The rule that uses and disclosures of patient information are limited to what the task actually requires.
- Workforce:
- Employees, volunteers, trainees and others under a facility's direct control, whether paid or not.
- Business associate:
- An outside person or company that handles patient information on a covered entity's behalf.
- Hostile work environment:
- A workplace made offensive by harassment that is severe or frequent, based on a protected trait.
- Mutual support:
- Team members checking each other's work and offering help when a teammate is overloaded.
- SBAR:
- A four-part reporting script: Situation, Background, Assessment, Recommendation.
- Handoff:
- Passing responsibility for work, and the information about it, from one person or shift to the next.
- Near miss:
- An event that did not injure a patient only because of luck or a timely catch.
- Culture of safety:
- A workplace where staff report errors without fear of blame so system causes can be fixed.
- Just culture:
- An approach that fixes system problems while still holding people accountable for risky or reckless choices.
- Continuing education (CE):
- Ongoing learning credits; CRCST renewal needs 12 technical, sterile processing credits each year.
- Suspension:
- The status after a missed renewal: not considered certified and not allowed to use the title.
- Grace period:
- The six weeks after a missed renewal date when late CE and fees can still be submitted.
Medical Terminology and Instruments
- -ectomy:
- Suffix meaning to cut out or surgically remove.
- -otomy:
- Suffix meaning a surgical incision or cutting into, without necessarily removing anything.
- -ostomy:
- Suffix meaning a surgically created opening from inside the body to the outside.
- -plasty:
- Suffix meaning surgical repair or reconstruction.
- -scopy:
- Suffix meaning examination of the inside of the body with a viewing instrument.
- Stoma:
- The opening on the skin created by an ostomy procedure.
- Supine:
- Lying flat on the back, face up.
- Prone:
- Lying flat on the front, face down.
- Proximal:
- Closer to the point where a limb attaches to the trunk.
- Distal:
- Farther from the point where a limb attaches to the trunk.
- Sagittal plane:
- A vertical plane that divides the body into right and left parts.
- Frontal (coronal) plane:
- A vertical plane that divides the body into front and back parts.
- Transverse plane:
- A horizontal plane that divides the body into upper and lower parts.
- Hemostat:
- A locking clamp with serrated jaws used to close off blood vessels and control bleeding.
- Box lock:
- The pinned hinge of a ringed instrument that keeps the jaw tips aligned.
- Ratchet:
- Interlocking teeth near the finger rings that hold an instrument's jaws closed.
- Serrations:
- Grooves cut across an instrument's jaws to help them grip tissue.
- Tungsten carbide insert:
- A very hard, wear-resistant jaw or blade piece, usually marked by gold handles.
- Self-retaining retractor:
- A tissue-spreading instrument that holds itself open with a ratchet or similar mechanism, needing no assistant.
- Rongeur:
- A sturdy bone-cutting instrument whose sharp, scoop-shaped tip bites off pieces of bone.