Free CRCST Preparation and Packaging practice test
HSPA gives this domain 21% of the exam. Answer each question, read why every choice is right or wrong, and see your score by topic.
- Instrument Inspection, Testing and Care
- Set Assembly and Chemical Indicators
- Packaging Methods and Labeling
CRCST Preparation and Packaging
A vascular set's tray list calls for scissors with rounded tips meant for delicate cutting and blunt dissection. Which scissors fit?
CRCST Preparation and Packaging practice questions and answers
Read each question, choose your answer, then open the explanation.
1.A vascular set's tray list calls for scissors with rounded tips meant for delicate cutting and blunt dissection. Which scissors fit?
- A.Mayo scissors
- B.Metzenbaum scissors
- C.Wire-cutting scissors
- D.Bandage scissors
Show answer
Answer: B. Metzenbaum scissors
Metzenbaum scissors are designed for delicate cutting and blunt dissection, and their rounded tips let the surgeon separate tissue without unintended sharp injury. Mayo scissors are heavier and are used on fascia and sutures.
Checked against Wikipedia, Metzenbaum scissors
2.While assembling a tray, a technician finds an extra instrument that is not on the tray list. What should the technician do?
- A.Leave it in the tray, since an extra instrument cannot harm anyone
- B.Add it to the tray list by hand so the counts will match in surgery
- C.Drop it into the nearest bin of spare instruments without telling anyone
- D.Report it to the supervisor and document it as a nonconformance
Show answer
Answer: D. Report it to the supervisor and document it as a nonconformance
The WHO manual says extra devices found during assembly should be reported to the supervisor and the nonconformance documented, just like missing devices. An unlisted instrument throws off the count and may belong to another set.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 (copy hosted by University of Washington)
3.What is the main purpose of checking each instrument against the set's tray list (count sheet)?
- A.To pick the sterilizer cycle that the set will need
- B.To record the name of the patient who will receive the set
- C.To work out how many months the set can stay on the shelf
- D.To confirm every listed item is present and correctly assembled
Show answer
Answer: D. To confirm every listed item is present and correctly assembled
The WHO manual says assembly ensures all devices are present according to the tray list and correctly assembled per the manufacturer's instructions.
4.How should hinged instruments such as hemostats be positioned in a set for sterilization?
- A.Open or unlocked
- B.Fully closed on the last ratchet
- C.Closed and taped shut
- D.Clamped together in pairs with rubber bands
Show answer
Answer: A. Open or unlocked
CDC says hinged instruments should be opened so the sterilant can reach all surfaces, including the box lock and jaws.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
5.How should a group of hemostats usually be organized in a tray?
- A.Loose in a pile at one end of the tray
- B.Stacked in a closed bundle held with tape
- C.Together on an instrument pin or stringer, open
- D.Sealed inside a peel pouch inside the tray
Show answer
Answer: C. Together on an instrument pin or stringer, open
The WHO manual suggests keeping similar devices together, such as artery forceps on a device pin, while keeping jointed instruments open.
6.How should a basin or other item with a concave surface be placed in a set?
- A.Tilted or on edge so water can drain
- B.Face up so condensate collects in the center
- C.Filled with a folded towel to soak up moisture
- D.Nested tightly inside another basin with no separation
Show answer
Answer: A. Tilted or on edge so water can drain
CDC's packaging guidance says to position items with concave surfaces so water drains off them, which helps prevent wet packs.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
7.Where should heavy instruments go when a set also has delicate instruments?
- A.On top of the delicate items to hold them in place
- B.Wrapped tightly together with the delicate items
- C.On the tray liner directly over the fine instruments
- D.Placed so they cannot damage delicate items, such as on the bottom
Show answer
Answer: D. Placed so they cannot damage delicate items, such as on the bottom
CDC and WHO both say heavy items should be positioned so they do not damage delicate items, which usually means on the bottom or in their own section.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
8.What is the purpose of a chemical indicator placed inside a package?
- A.To verify that sterilant penetrated to the inside
- B.To prove the contents are sterile
- C.To show the package left the sterilizer chamber at the end of the cycle
- D.To measure soil left after cleaning
Show answer
Answer: A. To verify that sterilant penetrated to the inside
CDC says an internal chemical indicator verifies sterilant penetration, while the external indicator shows the package went through a cycle. Neither proves sterility.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
9.When a set is opened, the internal chemical indicator has not changed as it should. What should happen?
- A.Use it if the external tape changed color
- B.Do not use the item; report and reprocess it
- C.Use it while the load BI is still pending
- D.Relabel the package with a new date and send it
Show answer
Answer: B. Do not use the item; report and reprocess it
CDC advises against using any item whose internal or external chemical indicator shows the process may have failed. The item is pulled, the failure reported, and the item reprocessed.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
10.What is the purpose of an ISO 11140 Type 1 (process) indicator, such as indicator tape?
- A.To tell processed packages apart from unprocessed ones
- B.To test the sterilizer air-removal function each morning before the first load
- C.To prove all critical cycle variables were met
- D.To replace the biological indicator for the load
Show answer
Answer: A. To tell processed packages apart from unprocessed ones
The WHO manual describes Type 1 process indicators as showing that a pack was exposed to the process, to distinguish processed from unprocessed units.
11.After each cleaning cycle, what should a technician check on instruments that have screws at their joints?
- A.That the screws are tight and have not loosened during cleaning
- B.That the screws have been removed so steam can reach the threads
- C.That each screw head has been oiled with a light machine oil
- D.That the screws have been marked with etched tray numbers
Show answer
Answer: A. That the screws are tight and have not loosened during cleaning
The WHO manual says jointed devices should be checked after each cleaning cycle to make sure their screws are tight and have not worked loose. A loose screw can keep the instrument from working as intended.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 (copy hosted by University of Washington)
12.During inspection of a cannulated orthopedic drill bit, what must the technician confirm about its channel?
- A.That it is capped at both ends to keep steam out of the lumen
- B.That it is patent, with nothing blocking it from end to end
- C.That it has been filled with lubricant to protect it during the cycle
- D.That it has dried with a thin protein film that resists corrosion
Show answer
Answer: B. That it is patent, with nothing blocking it from end to end
The WHO manual lists checking that cannulated devices have patent (open) channels as part of post-cleaning inspection. A blocked lumen may hide soil and keeps the sterilant from contacting the inner surface.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 (copy hosted by University of Washington)
13.Which pair of tools belongs at a post-cleaning inspection workstation in the prep and pack area?
- A.A lighted magnifier and an insulation tester for insulated devices
- B.An ultrasonic cleaner and an enzymatic soaking sink for any leftover soil
- C.A biological indicator incubator and a Bowie-Dick test pack
- D.A washer-disinfector loading rack and a container for used scalpel blades
Show answer
Answer: A. A lighted magnifier and an insulation tester for insulated devices
The inspection, assembly and packaging area needs good light with magnification to find soil and damage, and an insulation tester for insulated electrosurgical devices. Cleaning equipment stays in decontamination.
14.Why do departments use lighted magnification when inspecting instruments instead of relying on the naked eye?
- A.Magnification is needed mainly to read the catalog numbers stamped on instruments
- B.Soil in joints and grooves can be missed by an unaided visual check
- C.Magnified inspection makes later function testing of the instrument unnecessary
- D.Bright light helps instruments dry faster before they are wrapped and sterilized
Show answer
Answer: B. Soil in joints and grooves can be missed by an unaided visual check
In a study cited by CDC, most instruments that looked clean to the eye still had residual debris under magnification, especially in joints and grooves. Magnified, well-lit inspection improves the chance of catching that soil.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF
15.Which practice is expected before a technician starts inspecting and assembling clean instruments?
- A.Putting on the heavy-duty gloves used in decontamination
- B.Lubricating each instrument before checking it for soil
- C.Performing hand hygiene
- D.Placing the paper tray list inside the empty tray first
Show answer
Answer: C. Performing hand hygiene
The WHO manual lists hand hygiene before inspection work and keeping the workbench clean and disinfected between sessions, to avoid contaminating clean devices.
16.During a function check of a toothed clamp, what should the technician confirm about the jaws?
- A.The jaws are bent slightly apart so steam can get between them
- B.The jaws have been coated with a light machine oil to prevent rust
- C.The jaws stay locked on the last ratchet tooth while in storage
- D.The jaws and teeth line up when closed
Show answer
Answer: D. The jaws and teeth line up when closed
The WHO manual lists checking that jaws and teeth meet in line. Misaligned jaws do not grip properly and mean the instrument needs repair.
17.Which group of instruments should be checked for sharpness during inspection?
- A.Hemostats, needle holders and towel clamps
- B.Retractors, basins and vaginal specula
- C.Suction tips, cannulas and trocar sleeves
- D.Scissors, rongeurs, chisels and curettes
Show answer
Answer: D. Scissors, rongeurs, chisels and curettes
The WHO manual lists cutting instruments such as scissors, rongeurs, chisels and curettes as items whose edges must be checked for sharpness.
18.Why are insulated laparoscopic and electrosurgical instruments tested for insulation failure?
- A.Damaged insulation lets steam reach inner wiring and stops sterilization
- B.The test confirms the inside of the shaft was thoroughly cleaned
- C.Breaks in insulation can let current escape and burn the patient or staff
- D.Torn insulation makes the instrument heavier and harder for the surgeon to grip
Show answer
Answer: C. Breaks in insulation can let current escape and burn the patient or staff
Damaged insulation can let electrosurgical current leak to tissue or people, causing burns and fires. Many such burns go unnoticed during surgery, so instruments are tested before they are packaged.
Checked against Healthcare Purchasing News, Inspecting electrical instruments for insulation failure
19.A technician notices a crack at the box lock of a needle holder. What is the correct action?
- A.Remove it from the set, report it to the supervisor, and document it
- B.Put it back in the set, placed at the bottom so it is the last one picked up
- C.Throw it in the trash without telling anyone
- D.Bend the joint back into shape with pliers before packaging
Show answer
Answer: A. Remove it from the set, report it to the supervisor, and document it
The WHO manual says any damaged, incomplete or malfunctioning device should be reported to the supervisor immediately. It is removed from service and documented for repair.
20.Which type of lubricant is appropriate for surgical instruments that need it?
- A.Silicone spray from the maintenance shop
- B.Mineral oil or light machine oil
- C.Water-soluble instrument lubricant
- D.Petroleum jelly applied to the joints
Show answer
Answer: C. Water-soluble instrument lubricant
The WHO manual says lubricants should be water-soluble and used according to the device manufacturer's instructions; oily substances should not be used.
21.A technician is unsure which parts of a powered instrument to lubricate and how much to use. Where should they look?
- A.The device manufacturer's instructions for use
- B.The lubricant brand the operating room prefers
- C.Whatever amount the previous shift used
- D.A general rule of one drop per joint for all devices
Show answer
Answer: A. The device manufacturer's instructions for use
The WHO manual says the manufacturer's instructions govern which components need lubrication, which lubricant to use, and how to prepare and apply it, including amount and immersion time.
22.What gives stainless steel instruments their resistance to rust?
- A.A clear plastic coating applied at the factory
- B.A thin passive film of chromium and iron oxides
- C.A layer of lubricant baked on by repeated steam cycles
- D.Nickel plating that is renewed during each wash
Show answer
Answer: B. A thin passive film of chromium and iron oxides
A passive layer of chromium and iron oxides forms on stainless steel. When it is damaged and cannot re-form, the instrument becomes prone to corrosion.
Checked against Infection Control Today, Spotting, Staining, and Corrosion of Surgical Instruments
23.What must a sterilization packaging material do?
- A.Keep sterilant out so the contents stay dry
- B.Let sterilant in, then block microbes afterward
- C.Seal airtight so no air or steam can pass
- D.Be reusable for at least ten processing cycles
Show answer
Answer: B. Let sterilant in, then block microbes afterward
CDC says packaging must allow sterilant penetration, protect against contact contamination, act as a barrier to microbes and maintain sterility after processing.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
24.Which wrapping materials should not be used for hydrogen peroxide gas plasma sterilization?
- A.Polypropylene wraps
- B.Polyolefin pouches
- C.Wraps cleared for that system
- D.Linen and paper
Show answer
Answer: D. Linen and paper
CDC lists linens and paper as unacceptable for hydrogen peroxide gas plasma; the process requires synthetic packaging such as polypropylene wraps and polyolefin pouches.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
25.What does CDC recommend about packaging materials used in the department?
- A.They are the lowest-cost option the vendor offers
- B.They are the same type for every sterilization method
- C.They are made from reusable woven cotton only
- D.They are compatible with the process and FDA-cleared
Show answer
Answer: D. They are compatible with the process and FDA-cleared
CDC recommends that packaging materials be compatible with the sterilization process and have FDA 510(k) clearance.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF
26.What is sequential double wrapping?
- A.Two sheets laid together and wrapped as one, in a single pass
- B.Two sheets wrapped one after the other, a package within a package
- C.One sheet wrapped twice around the set with the same corners
- D.One sheet of wrap placed over a peel pouch and taped in place
Show answer
Answer: B. Two sheets wrapped one after the other, a package within a package
CDC describes sequential wrapping as using two sheets of wrap, one wrapped after the other, which creates a package within a package.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
27.What should be used to secure a wrapped pack closed?
- A.Safety pins through the wrap
- B.Paper clips at the corners
- C.Sterilization indicator tape
- D.Rubber bands around the pack
Show answer
Answer: C. Sterilization indicator tape
The WHO manual says wraps are secured with sterilization indicator tape, which also carries a chemical indicator.
28.What kind of item is best suited to a peel pouch?
- A.A heavy multilevel instrument set
- B.A stack of nested metal basins
- C.A single, lightweight device
- D.A full laparotomy tray
Show answer
Answer: C. A single, lightweight device
The WHO manual recommends pouches for single devices or light materials, packing one device per pouch.
29.Where should information be written on a paper-plastic peel pouch?
- A.On the paper side, right over the contents
- B.Inside the pouch on a slip of paper
- C.Across the heat seal itself
- D.On the clear plastic side, outside the seal
Show answer
Answer: D. On the clear plastic side, outside the seal
The WHO manual says pouches can be labeled outside the heat seal line on the clear laminate side, and warns not to write on the paper side.
30.Where should label information go on a wrapped instrument pack?
- A.On the indicator tape or a label
- B.Directly on the wrap with a ballpoint pen
- C.On a paper slip tucked under the wrap
- D.Inside the tray on the count sheet only
Show answer
Answer: A. On the indicator tape or a label
The WHO manual says label information should be put on sterilization indicator tape or a label, not on the packaging material.
31.What is event-related shelf life?
- A.The item stays sterile until an event such as a tear, wetting or a broken seal
- B.The item expires on a fixed date set by the sterilizer type used
- C.The item stays sterile forever, whatever happens to its package
- D.The item must be reprocessed after each scheduled inventory count, whatever its condition
Show answer
Answer: A. The item stays sterile until an event such as a tear, wetting or a broken seal
CDC explains that event-related practice treats a product as sterile until an event, such as a tear, wetting or a broken seal, causes contamination; handling raises the risk.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices