CRCST Packaging Methods and Labeling: study notes and practice questions

Part of the CRCST Preparation and Packaging domain, 21% of the exam. Packaging has to let sterilant in and then keep microbes out until the item is opened. Know which materials suit each sterilization method, how to wrap, pouch, seal and use rigid containers, how and where to label, and when a package must be treated as contaminated.

Study notes ↓Chapter quiz ↓All free questions ↓

All chapters

CRCST Packaging Methods and Labeling study notes

Packaging has to let sterilant in and then keep microbes out until the item is opened. Know which materials suit each sterilization method, how to wrap, pouch, seal and use rigid containers, how and where to label, and when a package must be treated as contaminated.

What packaging must do

  • Let the sterilant reach the contents, and let air out.
  • Act as a barrier to microbes once processing is over.
  • Protect the contents from contact contamination and damage while they are handled.
  • Keep the contents sterile until the package is opened, and allow them to be removed without contamination.
  • CDC says packaging should be compatible with the sterilization process and have FDA 510(k) clearance.
  • Choices include rigid containers, peel-open pouches, roll stock, and woven or nonwoven wraps; follow the packaging maker's instructions.
  • No single material suits every method, and the cheapest option is not automatically acceptable.

Matching packaging to the sterilization method

  • Steam: cellulose and nonwoven wraps, reusable woven fabric, peel pouches and rigid containers are all used.
  • Ethylene oxide (EtO): CDC rules out foil, polyvinyl chloride and polyvinylidene chloride (kitchen-type clear plastic wrap).
  • Hydrogen peroxide gas plasma: do not use linen or paper.
  • Paper and other cellulose materials soak up hydrogen peroxide vapor, which interferes with forming the plasma during the cycle.
  • Plasma needs fully synthetic packaging, such as synthetic nonwoven wraps and non-cellulose pouches like Tyvek-type polyolefin.
  • Liquid peracetic acid: porous packaging would become soaked, so items are run unwrapped in a machine-specific cassette and used right away, not stored.

Wrapping a set

  • Sequential wrapping: two sheets wrapped one after the other, making a package within a package.
  • Simultaneous (nonsequential) wrapping: two sheets laid together and wrapped in one pass, which saves time but still gives two layers.
  • Wrap snugly; tenting, gaps, billowing and air pockets can compromise sterility, so rewrap a loose pack rather than taping over gaps.
  • Envelope fold: place the set diagonally, slightly off center, and fold the corners over it, leaving a tab for aseptic opening.
  • Square (parcel) fold: center the set with the tray's long edge parallel to the wrap's long edge.
  • Close the pack with sterilization indicator tape, never pins, paper clips or rubber bands.
  • Before each reuse, inspect woven fabric wraps for holes and thin, threadbare areas, which make them ineffective.
  • Woven fabric is a weaker microbial barrier than many single-use wraps and is not used as a single layer.

Peel pouches and heat sealing

  • Pouches are for single, lightweight devices, one device per pouch; heavy sets, basins and large trays are wrapped or containerized.
  • Smooth out creases at the seal area first, because creases cause an incomplete or uneven seal.
  • Press out as much air as possible before sealing, then check that the seal is complete with no open spots or bubbles.
  • Double pouching: pick an inner pouch one size smaller, do not fold its edges, and place paper against paper.
  • Folded edges on an inner pouch can trap air that blocks sterilization.
  • Never staple a pouch, because staples punch holes in it; rubber bands and glue are not acceptable either.
  • If the sealer breaks, the WHO manual describes folding the open end and securing the folds with tape; self-seal pouches are used per their maker's instructions.
  • Seal a test piece of each packaging type on each sealer every day and check the seal's integrity and strength.
  • Heat sealers also need full servicing, including temperature calibration, at least once every 12 months.

Rigid containers

  • Holes in the lid and base are lined with a filter that lets steam through but keeps microbes out afterward.
  • Before use, check the filters, seals and tamper-evident devices, and look for any damage.
  • Load containers with sensible density to avoid moisture problems and long drying times.
  • After use, take containers apart, wash and dry them; they need routine inspection and maintenance.

Labeling

  • Label packages before they are sterilized.
  • CDC: the load number should identify which sterilizer ran it, the cycle or load number, the date it was sterilized and, if used, the expiration date.
  • The WHO manual adds the product name and who wrapped it; labels identify the processing, not the patient.
  • On wrapped packs, write on the indicator tape or a label, not on the wrap itself.
  • On paper-plastic pouches, write on the clear plastic side, outside the heat seal; ink can soak through the paper side.
  • Use an indelible, non-bleeding, non-toxic marker; sharp-tipped, water-based and ballpoint pens can damage the package.
  • Never put a paper slip inside a package, since it sheds lint and cannot be read from outside.
  • A peel-off (piggyback) batch label or computer system lets operating room staff put the lot information in the patient's record, which makes recalls possible.

Shelf life and package integrity

  • Event-related shelf life: an item stays sterile until something happens to it, such as a tear, wetting or a broken seal.
  • The more a package is handled, the more likely it is to become contaminated.
  • A package that is wet, torn or punctured, or past any expiration date, is not used.
  • A wet package is contaminated, because moisture can carry microbes in; it is reprocessed, and letting it dry on the shelf does not fix it.
  • A dropped package must be inspected for damage.
  • If it is sealed by heat in impervious plastic and the seal has not broken, CDC treats it as not contaminated, and undamaged plastic-packaged items need not be reprocessed.

Key terms

Sterile barrier
Packaging that lets sterilant in, then blocks microbes so contents stay sterile until opened.
Sequential wrapping
Two sheets wrapped one after the other, creating a package within a package.
Simultaneous wrapping
Two sheets laid together and folded as one, giving two layers with a single wrapping step.
Envelope fold
A wrapping method with the set placed diagonally, slightly off center, and the corners folded over it.
Square fold
A wrapping method, also called parcel fold, with the set centered and parallel to the wrap's long edge.
Tenting
Loose wrap standing away from the contents, leaving air pockets that can compromise the barrier.
Peel pouch
A heat-sealed paper-plastic or synthetic package for a single, lightweight device.
Heat sealer
A machine that closes pouches with heat and pressure; tested daily and calibrated at least yearly.
Rigid container
A reusable metal or polymer box with filtered vents, used mainly for large steam-sterilized sets.
Tyvek
A synthetic polyolefin material with low chemical absorption, suitable for hydrogen peroxide plasma packaging.
Load number
Label code identifying the sterilizer, the cycle or load, the date and any expiration date.
Event-related shelf life
Treating an item as sterile until damage, moisture, a broken seal or other event compromises it.
Batch control label
A peel-off label recording lot details, placed in the patient's record to allow tracing and recall.
Indicator tape
Adhesive tape that closes wrapped packs and carries a process indicator that changes color in a cycle.

Common mistakes

  • Picking a wrap that keeps sterilant out or seals airtight: packaging must let sterilant in and then act as a microbial barrier.
  • Using paper or linen in hydrogen peroxide gas plasma, or foil and kitchen plastic wrap in EtO: match the material to the method.
  • Stapling a pouch, banding it, or adding tape over staples when the sealer is broken: staples perforate the barrier; use an approved method.
  • Writing on the paper side of a pouch or directly on the wrap with a ballpoint: write on the plastic side outside the seal, or on tape or a label, with an indelible marker.
  • Folding a same-size inner pouch to fit, or facing plastic to paper: use a smaller unfolded inner pouch with paper facing paper.
  • Assuming any dropped or damp package is fine once it dries, or that any dropped package is ruined: wet means contaminated, but an intact heat-sealed impervious plastic package is not.

Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities: Sterilizing Practices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), copy hosted by University of Washington

Chapter quiz

6 questions from this chapter, with an explanation after each answer.

1/6

Packaging Methods and Labeling

What must a sterilization packaging material do?

Packaging Methods and Labeling questions and answers

Read each question, choose your answer, then open the explanation.

  1. 1.What must a sterilization packaging material do?

    • A.Keep sterilant out so the contents stay dry
    • B.Let sterilant in, then block microbes afterward
    • C.Seal airtight so no air or steam can pass
    • D.Be reusable for at least ten processing cycles
    Show answer

    Answer: B. Let sterilant in, then block microbes afterward

    CDC says packaging must allow sterilant penetration, protect against contact contamination, act as a barrier to microbes and maintain sterility after processing.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  2. 2.Which wrapping materials should not be used for hydrogen peroxide gas plasma sterilization?

    • A.Polypropylene wraps
    • B.Polyolefin pouches
    • C.Wraps cleared for that system
    • D.Linen and paper
    Show answer

    Answer: D. Linen and paper

    CDC lists linens and paper as unacceptable for hydrogen peroxide gas plasma; the process requires synthetic packaging such as polypropylene wraps and polyolefin pouches.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  3. 3.What does CDC recommend about packaging materials used in the department?

    • A.They are the lowest-cost option the vendor offers
    • B.They are the same type for every sterilization method
    • C.They are made from reusable woven cotton only
    • D.They are compatible with the process and FDA-cleared
    Show answer

    Answer: D. They are compatible with the process and FDA-cleared

    CDC recommends that packaging materials be compatible with the sterilization process and have FDA 510(k) clearance.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF

  4. 4.What is sequential double wrapping?

    • A.Two sheets laid together and wrapped as one, in a single pass
    • B.Two sheets wrapped one after the other, a package within a package
    • C.One sheet wrapped twice around the set with the same corners
    • D.One sheet of wrap placed over a peel pouch and taped in place
    Show answer

    Answer: B. Two sheets wrapped one after the other, a package within a package

    CDC describes sequential wrapping as using two sheets of wrap, one wrapped after the other, which creates a package within a package.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  5. 5.What should be used to secure a wrapped pack closed?

    • A.Safety pins through the wrap
    • B.Paper clips at the corners
    • C.Sterilization indicator tape
    • D.Rubber bands around the pack
    Show answer

    Answer: C. Sterilization indicator tape

    The WHO manual says wraps are secured with sterilization indicator tape, which also carries a chemical indicator.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  6. 6.What kind of item is best suited to a peel pouch?

    • A.A heavy multilevel instrument set
    • B.A stack of nested metal basins
    • C.A single, lightweight device
    • D.A full laparotomy tray
    Show answer

    Answer: C. A single, lightweight device

    The WHO manual recommends pouches for single devices or light materials, packing one device per pouch.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  7. 7.Where should information be written on a paper-plastic peel pouch?

    • A.On the paper side, right over the contents
    • B.Inside the pouch on a slip of paper
    • C.Across the heat seal itself
    • D.On the clear plastic side, outside the seal
    Show answer

    Answer: D. On the clear plastic side, outside the seal

    The WHO manual says pouches can be labeled outside the heat seal line on the clear laminate side, and warns not to write on the paper side.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  8. 8.Where should label information go on a wrapped instrument pack?

    • A.On the indicator tape or a label
    • B.Directly on the wrap with a ballpoint pen
    • C.On a paper slip tucked under the wrap
    • D.Inside the tray on the count sheet only
    Show answer

    Answer: A. On the indicator tape or a label

    The WHO manual says label information should be put on sterilization indicator tape or a label, not on the packaging material.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  9. 9.What is event-related shelf life?

    • A.The item stays sterile until an event such as a tear, wetting or a broken seal
    • B.The item expires on a fixed date set by the sterilizer type used
    • C.The item stays sterile forever, whatever happens to its package
    • D.The item must be reprocessed after each scheduled inventory count, whatever its condition
    Show answer

    Answer: A. The item stays sterile until an event such as a tear, wetting or a broken seal

    CDC explains that event-related practice treats a product as sterile until an event, such as a tear, wetting or a broken seal, causes contamination; handling raises the risk.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices