Assembly puts the right instruments into the right tray, arranged so sterilant reaches every surface and the set dries. Know how to work from the tray list, how to position instruments, what internal and external chemical indicators show, the six indicator types, and the extra rules for implants and loaner sets.
Working from the tray list
- Assembly makes sure every device on the tray list is present, put together correctly per the maker's instructions for use (IFU), and placed so the user can work easily.
- Check each device against the list written for that specific tray.
- When two instruments look alike, match the exact item the list names (for example its catalog number and size) instead of guessing by eye.
- If a listed item is missing, report it to the supervisor and document it; report extra items found in a tray the same way.
- Do not quietly swap in an item from another set, send the set short without notice, or hold it back without telling anyone.
- Keep the paper tray list outside the instrument tray, because paper sheds lint that can reach instruments and harm the patient.
- Where tracking is done by hand, the devices in a set are documented on its tray list.
Positioning instruments for sterilant contact
- Sterilant must touch every surface, so hinged instruments such as hemostats go in open or unlocked.
- If a ratcheted instrument has to be closed, close it on the first ratchet only.
- Never hold instruments shut with tape or rubber bands, and never close them on the last ratchet.
- Keep similar instruments together, for example hemostats open on an instrument pin or stringer.
- Take apart devices with removable parts unless the device maker's instructions say otherwise.
- After taking a device apart, keep all its parts together in the same set so it can be put back together quickly.
- Use perforated trays, put a tray liner on the bottom, and lay instruments in a single layer where you can.
Weight, moisture and protecting delicate items
- Place basins and other concave items tilted or on edge so water drains out.
- Separate nested hollowware with nonporous material and face all openings the same way.
- Put heavy instruments where they cannot crush delicate ones, often on the bottom.
- Spread instruments evenly by weight across the tray so condensate does not run together in one spot.
- Spread plastic items evenly through the tray instead of grouping them in one section.
- Protect sharp tips with protectors validated for steam penetration, fitted so they are not too tight.
- CDC no longer gives a set weight limit; base weight on the set's design, how dense the instruments are, and how the metal mass is distributed.
- Heavy metal mass is one cause of wet packs, meaning moisture left inside the case or tray after the cycle.
Internal and external chemical indicators
- An external indicator, such as indicator tape, goes on the outside of each pack and shows only that the pack went through a cycle.
- An internal indicator goes inside the pack to show that sterilant got in.
- Because its job is to check penetration, put the internal indicator where sterilant has the hardest path, following the set and indicator makers' directions.
- CDC calls for internal and external indicators on each load, but if the internal one can be seen through the package, as in a peel pouch, no external one is needed.
- No chemical indicator proves an item is sterile; only a biological indicator measures whether microbes were killed.
- Chemical indicators are used along with biological indicators, never in place of them.
- If an internal or external indicator suggests the item was not properly processed, do not use it; report it and send it back for reprocessing.
The six chemical indicator types
International standard ISO 11140 sorts chemical indicators into six types. Older CDC text calls them classes and lists five.
- Type 1, process indicator: on packs or containers, it tells processed units from unprocessed ones (indicator tape is an example).
- Type 2, specific-test indicator: used in special test procedures such as the Bowie-Dick air-removal test.
- Type 3, single-variable indicator: reacts to one critical variable, such as temperature.
- Type 4, multivariable indicator: reacts to two or more critical variables, such as time and temperature.
- Type 5, integrating indicator: reacts to all critical variables and is meant to match or beat the performance requirements for biological indicators.
- Type 6, emulating indicator: reacts to all critical variables and is built to match those of a specified cycle.
The Bowie-Dick test
- It checks air removal and air leaks in a dynamic-air-removal (prevacuum) steam sterilizer.
- Run it each day the sterilizer is used, in an otherwise empty chamber, before the first load.
- A pass is an even color change across the test sheet; trapped air leaves a spot where steam could not reach.
- If it fails, do not use the sterilizer until it has been inspected and passes the test.
- It is not a test of sterility and does not replace a biological indicator.
Implants and loaner sets
- Run a biological indicator with every load that contains implantable items.
- Quarantine implants, whenever possible, until that biological indicator reads negative.
- The WHO manual says immediate-use (flash) steam sterilization must never be used for implants; CDC does not recommend it and allows it only when unavoidable, with full records.
- A reusable device can be reused safely only when its maker's reprocessing instructions are followed completely after every use.
- So a loaner set marked as already sterilized by a vendor is still checked against the vendor's list, then cleaned, inspected and processed per the IFU before use.
Key terms
- Tray list
- The count sheet naming every device a set should contain, used to check completeness during assembly.
- Stringer
- A pin or rack that holds ringed instruments together, open, inside a tray.
- First ratchet
- The lightest locked position on a ratcheted instrument, the only closed position allowed in a set.
- Tip protector
- A cover over a sharp end, validated for steam penetration and fitted loosely enough for sterilant to reach.
- Wet pack
- A package with moisture inside the case or tray after the cycle ends; its contents are considered contaminated.
- External chemical indicator
- A marker on the outside of a package showing it went through a cycle, not that it is sterile.
- Internal chemical indicator
- A marker placed inside a package to show the sterilant reached the contents.
- Process indicator
- ISO 11140 Type 1; shows a pack was exposed to the process, separating processed from unprocessed units.
- Integrating indicator
- ISO 11140 Type 5; reacts to all critical variables and is meant to equal or exceed biological indicator performance requirements.
- Emulating indicator
- ISO 11140 Type 6; reacts to all critical variables and matches those of a specified cycle.
- Bowie-Dick test
- A daily empty-chamber check of air removal and air leaks in a prevacuum steam sterilizer.
- Biological indicator
- A test containing resistant spores; the only monitor that measures whether a cycle killed microbes.
- Quarantine
- Holding processed items back from use until a pending test result, such as a spore test, comes back.
- Loaner set
- Instruments borrowed from a vendor or another facility for a procedure, processed here per the maker's instructions before use.
- Immediate-use steam sterilization
- A shortened steam cycle for items needed at once; the WHO manual says never to use it for implants.
Common mistakes
- Reading color-changed tape as proof the contents are sterile: an external indicator only shows the pack went through a cycle.
- Using a set because the tape changed even though the internal indicator did not: a failed indicator means the item is not used.
- Putting the internal indicator in the easiest spot, such as on top or in an open corner: it belongs where sterilant has the hardest path.
- Closing hemostats on the last ratchet, taping them, or banding them together: hinged instruments go in open, or on the first ratchet at most.
- Placing the paper tray list inside the tray or substituting a missing instrument without noting it: lists stay outside, and shortages are reported and documented.
- Mixing up indicator types: Type 1 is tape on packs, Type 2 is for the Bowie-Dick test, Type 5 is tied to biological indicator performance, and Type 6 matches a specific cycle.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities: Sterilizing Practices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), copy hosted by University of Washington; FDA, Reprocessing of Reusable Medical Devices