After cleaning, every instrument is checked for leftover soil, tested to make sure it works, and cared for so it lasts. Know what to look at, what each function check proves, what to do with a damaged item, what causes stains and corrosion, and how lubrication fits in.
Setting up the inspection station
- Inspection happens on the clean side, after items are cleaned and dried, in the same area where sets are assembled and packaged.
- The bench needs bright light with a magnifier or magnifying lamp, a light source, and an insulation tester for insulated electrosurgical devices.
- Cleaning tools such as sinks, ultrasonic cleaners and washer racks stay in decontamination; a soiled item goes back there instead.
- Do hand hygiene before you start, and keep the bench tidy and disinfected between work sessions.
- Clean-side attire is a hair cover and a clean, low-lint uniform; heavy-duty decontamination gloves are not brought to this area.
- Inspect one set at a time so items from different sets do not get mixed up.
Looking for retained soil
An instrument that looks clean may not be clean. CDC describes a study in which 91% of instruments passed a visual check, yet 84% showed debris when examined under a microscope.
- In that study, debris hid where the insulating sheath of a laparoscopic instrument meets its activating mechanism, and in the joints and grooves of forceps.
- Give box joints, jaw serrations and other crevices the closest look, because brushes reach them poorly.
- Smooth, open surfaces such as finger rings and flat shanks rarely hide soil.
- Leftover blood or tissue can shield microbes from disinfection or sterilization, according to FDA.
- Any item with visible soil goes back to decontamination to be cleaned again.
Function checks
- Hinges and box locks should open and close smoothly without sticking.
- Jaws and teeth should meet evenly when the instrument is closed.
- Ratchets should engage and hold; one that springs open has failed.
- Cutting edges on scissors, rongeurs, chisels and curettes are checked for sharpness.
- Use the sharpness test method and test material the instrument maker's instructions for use (IFU) call for; there is no single test medium for every scissor.
- Hollow (cannulated) devices are checked to confirm their channels are open.
- Multi-part instruments are put together to confirm every part is present and works.
- Telescopes and light cables are both function-checked the way their makers' IFUs describe.
- Every instrument is checked on every pass through reprocessing; passing last time does not count.
Insulated electrosurgical instruments
- A break in the insulation can let electrical current reach the patient, staff or flammable materials, causing burns or fire.
- Many stray burns during laparoscopic surgery go unnoticed while the case is under way.
- Tiny defects can be invisible, and a visual check cannot find every path current might leak through.
- So insulated items get a visual check plus a test with equipment built to detect insulation failure.
- Test robotic and bipolar instruments and their cords too, not only laparoscopic ones.
- FDA's advice is simple: if a defect is found, do not use the device.
When an instrument is damaged
- Report any damaged, incomplete or malfunctioning device to the supervisor right away.
- Take it out of the set, tag it as defective, and document it so it can be repaired or replaced.
- Damaged surfaces let dirt and bacteria build up and can be dangerous to patients and staff.
- A crack at a box lock is a safety problem, not a cosmetic one, even if the instrument still opens and closes.
- Do not fix instruments at the bench by bending, filing or taping them; repairs go to a qualified repair service.
- Never send a known defective device to the operating room with a warning note instead.
Stainless steel, stains and corrosion
- Stainless steel resists rust because of an invisible surface film made mostly of chromium and iron oxides.
- That film is harmed by scratches, harsh acid or alkaline cleaners, abrasive pads and brushes, and dried soil.
- Chlorine bleach and saline are among the worst exposures; do not soak instruments in saline.
- Pitting shows as small black holes caused by blood, chlorides or other corrosive solutions; remove and replace the instrument and work with the repair company to find the cause.
- Etching removes the protective coating and promotes corrosion, so check with the instrument maker before etching marks onto instruments.
- Spots sit on the surface and wipe off; white, chalky spots usually come from minerals in hard rinse water, and a deionized or softened final rinse helps prevent them.
- A rainbow, gold or rusty tint can appear when different metals share one solution, because metal moves from one to the other; wash them apart or use a neutral cleaner safe for soft metals.
- Stains or rust-colored deposits that appear only after sterilizing in reusable woven wraps can come from detergent, alkaline or acidic residue left in the fabric by laundering; a pH test of the wraps and a review of the laundry process are the fix.
Lubrication
- A lubricant is there only to protect the device, and it must dissolve in water.
- Never use oily products such as mineral oil, machine oil or petroleum jelly.
- Lubricate only after the item is clean and free of visible soil and rust, and before it is sterilized.
- The device IFU says which parts need lubricant, which product to use, how to mix and apply it, how much, and any soak time.
- A lubricant that does not suit the device or process can stop sterilization from working, create harmful by-products, and harm the device or even the sterilizer.
- If a clean joint feels stiff, recheck it for soil or damage first, then lubricate as the IFU directs; do not force it open and shut.
- Throw out lubricant that is past its expiry date or looks dirty.
- Regular use of a water-soluble lubricant also helps support the protective surface film.
Key terms
- Box lock
- The hinged joint where the two halves of a clamp or hemostat cross; a common place for hidden soil and cracks.
- Serrations
- Grooves or ridges cut into instrument jaws to improve grip; they trap debris and need close inspection.
- Ratchet
- Interlocking teeth near the finger rings that hold an instrument's jaws closed at set positions.
- Function check
- Testing that an instrument works as intended, such as jaw alignment, ratchet hold, smooth hinges and sharp edges.
- Retained soil
- Blood, tissue or other debris left on a device after cleaning, which can shield microbes from sterilization.
- Insulation tester
- A device that detects breaks in the coating of electrosurgical instruments that could let current escape.
- Cannulated device
- An instrument with a hollow channel, such as a suction tip, checked to confirm the channel is open.
- Passive layer
- The thin chromium and iron oxide film on stainless steel that protects it from rust.
- Pitting
- Small dark holes where metal has been lost to corrosion, often from blood, chlorides or harsh solutions.
- Spotting
- Loose deposits sitting on an instrument's surface, often minerals from hard water, that can be wiped away.
- Galvanic corrosion
- Metal transferring from one metal to another when dissimilar metals share a solution, leaving rainbow, gold or rusty tints.
- Water-soluble lubricant
- An instrument lubricant that dissolves in water, so it does not block sterilant from reaching the metal.
- Instructions for use (IFU)
- The manufacturer's written directions for cleaning, testing, lubricating and sterilizing a specific device.
- Deionized water
- Water with dissolved minerals removed, used for final rinsing to prevent spots and deposits.
Common mistakes
- Wiping dried blood off a box lock with an alcohol pad or a dry brush at the prep table: visibly soiled items go back to decontamination to be recleaned.
- Trusting the sterilizer to deal with leftover soil: residue can shield microbes, so the item may not be sterilized.
- Thinking a magnified check replaces function testing: cleanliness and function are separate checks, and both are needed.
- Taping over, retesting or sending a note with an instrument that failed an insulation test: a defective device is removed from service and reported.
- Using mineral oil, machine oil, petroleum jelly or a shop spray on stiff joints: only the water-soluble lubricant the IFU names, and only after cleaning.
- Buffing pits with an abrasive pad or calling pits a stain: pitting is metal loss, and abrasives make corrosion worse.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), copy hosted by University of Washington; FDA, Reprocessing of Reusable Medical Devices; Healthcare Purchasing News, Inspecting electrical instruments for insulation failure; Healthcare Purchasing News, Understanding your P's, Q's, and S's in instrument care & handling; Infection Control Today, Spotting, Staining, and Corrosion of Surgical Instruments