CRCST Instrument Inspection, Testing and Care: study notes and practice questions

Part of the CRCST Preparation and Packaging domain, 21% of the exam. After cleaning, every instrument is checked for leftover soil, tested to make sure it works, and cared for so it lasts. Know what to look at, what each function check proves, what to do with a damaged item, what causes stains and corrosion, and how lubrication fits in.

Study notes ↓Chapter quiz ↓All free questions ↓

All chapters

CRCST Instrument Inspection, Testing and Care study notes

After cleaning, every instrument is checked for leftover soil, tested to make sure it works, and cared for so it lasts. Know what to look at, what each function check proves, what to do with a damaged item, what causes stains and corrosion, and how lubrication fits in.

Setting up the inspection station

  • Inspection happens on the clean side, after items are cleaned and dried, in the same area where sets are assembled and packaged.
  • The bench needs bright light with a magnifier or magnifying lamp, a light source, and an insulation tester for insulated electrosurgical devices.
  • Cleaning tools such as sinks, ultrasonic cleaners and washer racks stay in decontamination; a soiled item goes back there instead.
  • Do hand hygiene before you start, and keep the bench tidy and disinfected between work sessions.
  • Clean-side attire is a hair cover and a clean, low-lint uniform; heavy-duty decontamination gloves are not brought to this area.
  • Inspect one set at a time so items from different sets do not get mixed up.

Looking for retained soil

An instrument that looks clean may not be clean. CDC describes a study in which 91% of instruments passed a visual check, yet 84% showed debris when examined under a microscope.

  • In that study, debris hid where the insulating sheath of a laparoscopic instrument meets its activating mechanism, and in the joints and grooves of forceps.
  • Give box joints, jaw serrations and other crevices the closest look, because brushes reach them poorly.
  • Smooth, open surfaces such as finger rings and flat shanks rarely hide soil.
  • Leftover blood or tissue can shield microbes from disinfection or sterilization, according to FDA.
  • Any item with visible soil goes back to decontamination to be cleaned again.

Function checks

  • Hinges and box locks should open and close smoothly without sticking.
  • Jaws and teeth should meet evenly when the instrument is closed.
  • Ratchets should engage and hold; one that springs open has failed.
  • Cutting edges on scissors, rongeurs, chisels and curettes are checked for sharpness.
  • Use the sharpness test method and test material the instrument maker's instructions for use (IFU) call for; there is no single test medium for every scissor.
  • Hollow (cannulated) devices are checked to confirm their channels are open.
  • Multi-part instruments are put together to confirm every part is present and works.
  • Telescopes and light cables are both function-checked the way their makers' IFUs describe.
  • Every instrument is checked on every pass through reprocessing; passing last time does not count.

Insulated electrosurgical instruments

  • A break in the insulation can let electrical current reach the patient, staff or flammable materials, causing burns or fire.
  • Many stray burns during laparoscopic surgery go unnoticed while the case is under way.
  • Tiny defects can be invisible, and a visual check cannot find every path current might leak through.
  • So insulated items get a visual check plus a test with equipment built to detect insulation failure.
  • Test robotic and bipolar instruments and their cords too, not only laparoscopic ones.
  • FDA's advice is simple: if a defect is found, do not use the device.

When an instrument is damaged

  • Report any damaged, incomplete or malfunctioning device to the supervisor right away.
  • Take it out of the set, tag it as defective, and document it so it can be repaired or replaced.
  • Damaged surfaces let dirt and bacteria build up and can be dangerous to patients and staff.
  • A crack at a box lock is a safety problem, not a cosmetic one, even if the instrument still opens and closes.
  • Do not fix instruments at the bench by bending, filing or taping them; repairs go to a qualified repair service.
  • Never send a known defective device to the operating room with a warning note instead.

Stainless steel, stains and corrosion

  • Stainless steel resists rust because of an invisible surface film made mostly of chromium and iron oxides.
  • That film is harmed by scratches, harsh acid or alkaline cleaners, abrasive pads and brushes, and dried soil.
  • Chlorine bleach and saline are among the worst exposures; do not soak instruments in saline.
  • Pitting shows as small black holes caused by blood, chlorides or other corrosive solutions; remove and replace the instrument and work with the repair company to find the cause.
  • Etching removes the protective coating and promotes corrosion, so check with the instrument maker before etching marks onto instruments.
  • Spots sit on the surface and wipe off; white, chalky spots usually come from minerals in hard rinse water, and a deionized or softened final rinse helps prevent them.
  • A rainbow, gold or rusty tint can appear when different metals share one solution, because metal moves from one to the other; wash them apart or use a neutral cleaner safe for soft metals.
  • Stains or rust-colored deposits that appear only after sterilizing in reusable woven wraps can come from detergent, alkaline or acidic residue left in the fabric by laundering; a pH test of the wraps and a review of the laundry process are the fix.

Lubrication

  • A lubricant is there only to protect the device, and it must dissolve in water.
  • Never use oily products such as mineral oil, machine oil or petroleum jelly.
  • Lubricate only after the item is clean and free of visible soil and rust, and before it is sterilized.
  • The device IFU says which parts need lubricant, which product to use, how to mix and apply it, how much, and any soak time.
  • A lubricant that does not suit the device or process can stop sterilization from working, create harmful by-products, and harm the device or even the sterilizer.
  • If a clean joint feels stiff, recheck it for soil or damage first, then lubricate as the IFU directs; do not force it open and shut.
  • Throw out lubricant that is past its expiry date or looks dirty.
  • Regular use of a water-soluble lubricant also helps support the protective surface film.

Key terms

Box lock
The hinged joint where the two halves of a clamp or hemostat cross; a common place for hidden soil and cracks.
Serrations
Grooves or ridges cut into instrument jaws to improve grip; they trap debris and need close inspection.
Ratchet
Interlocking teeth near the finger rings that hold an instrument's jaws closed at set positions.
Function check
Testing that an instrument works as intended, such as jaw alignment, ratchet hold, smooth hinges and sharp edges.
Retained soil
Blood, tissue or other debris left on a device after cleaning, which can shield microbes from sterilization.
Insulation tester
A device that detects breaks in the coating of electrosurgical instruments that could let current escape.
Cannulated device
An instrument with a hollow channel, such as a suction tip, checked to confirm the channel is open.
Passive layer
The thin chromium and iron oxide film on stainless steel that protects it from rust.
Pitting
Small dark holes where metal has been lost to corrosion, often from blood, chlorides or harsh solutions.
Spotting
Loose deposits sitting on an instrument's surface, often minerals from hard water, that can be wiped away.
Galvanic corrosion
Metal transferring from one metal to another when dissimilar metals share a solution, leaving rainbow, gold or rusty tints.
Water-soluble lubricant
An instrument lubricant that dissolves in water, so it does not block sterilant from reaching the metal.
Instructions for use (IFU)
The manufacturer's written directions for cleaning, testing, lubricating and sterilizing a specific device.
Deionized water
Water with dissolved minerals removed, used for final rinsing to prevent spots and deposits.

Common mistakes

  • Wiping dried blood off a box lock with an alcohol pad or a dry brush at the prep table: visibly soiled items go back to decontamination to be recleaned.
  • Trusting the sterilizer to deal with leftover soil: residue can shield microbes, so the item may not be sterilized.
  • Thinking a magnified check replaces function testing: cleanliness and function are separate checks, and both are needed.
  • Taping over, retesting or sending a note with an instrument that failed an insulation test: a defective device is removed from service and reported.
  • Using mineral oil, machine oil, petroleum jelly or a shop spray on stiff joints: only the water-soluble lubricant the IFU names, and only after cleaning.
  • Buffing pits with an abrasive pad or calling pits a stain: pitting is metal loss, and abrasives make corrosion worse.

Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), copy hosted by University of Washington; FDA, Reprocessing of Reusable Medical Devices; Healthcare Purchasing News, Inspecting electrical instruments for insulation failure; Healthcare Purchasing News, Understanding your P's, Q's, and S's in instrument care & handling; Infection Control Today, Spotting, Staining, and Corrosion of Surgical Instruments

Chapter quiz

6 questions from this chapter, with an explanation after each answer.

1/6

Instrument Inspection, Testing and Care

After each cleaning cycle, what should a technician check on instruments that have screws at their joints?

Instrument Inspection, Testing and Care questions and answers

Read each question, choose your answer, then open the explanation.

  1. 1.After each cleaning cycle, what should a technician check on instruments that have screws at their joints?

    • A.That the screws are tight and have not loosened during cleaning
    • B.That the screws have been removed so steam can reach the threads
    • C.That each screw head has been oiled with a light machine oil
    • D.That the screws have been marked with etched tray numbers
    Show answer

    Answer: A. That the screws are tight and have not loosened during cleaning

    The WHO manual says jointed devices should be checked after each cleaning cycle to make sure their screws are tight and have not worked loose. A loose screw can keep the instrument from working as intended.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 (copy hosted by University of Washington)

  2. 2.During inspection of a cannulated orthopedic drill bit, what must the technician confirm about its channel?

    • A.That it is capped at both ends to keep steam out of the lumen
    • B.That it is patent, with nothing blocking it from end to end
    • C.That it has been filled with lubricant to protect it during the cycle
    • D.That it has dried with a thin protein film that resists corrosion
    Show answer

    Answer: B. That it is patent, with nothing blocking it from end to end

    The WHO manual lists checking that cannulated devices have patent (open) channels as part of post-cleaning inspection. A blocked lumen may hide soil and keeps the sterilant from contacting the inner surface.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 (copy hosted by University of Washington)

  3. 3.Which pair of tools belongs at a post-cleaning inspection workstation in the prep and pack area?

    • A.A lighted magnifier and an insulation tester for insulated devices
    • B.An ultrasonic cleaner and an enzymatic soaking sink for any leftover soil
    • C.A biological indicator incubator and a Bowie-Dick test pack
    • D.A washer-disinfector loading rack and a container for used scalpel blades
    Show answer

    Answer: A. A lighted magnifier and an insulation tester for insulated devices

    The inspection, assembly and packaging area needs good light with magnification to find soil and damage, and an insulation tester for insulated electrosurgical devices. Cleaning equipment stays in decontamination.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  4. 4.Why do departments use lighted magnification when inspecting instruments instead of relying on the naked eye?

    • A.Magnification is needed mainly to read the catalog numbers stamped on instruments
    • B.Soil in joints and grooves can be missed by an unaided visual check
    • C.Magnified inspection makes later function testing of the instrument unnecessary
    • D.Bright light helps instruments dry faster before they are wrapped and sterilized
    Show answer

    Answer: B. Soil in joints and grooves can be missed by an unaided visual check

    In a study cited by CDC, most instruments that looked clean to the eye still had residual debris under magnification, especially in joints and grooves. Magnified, well-lit inspection improves the chance of catching that soil.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last update June 2024), full PDF

  5. 5.Which practice is expected before a technician starts inspecting and assembling clean instruments?

    • A.Putting on the heavy-duty gloves used in decontamination
    • B.Lubricating each instrument before checking it for soil
    • C.Performing hand hygiene
    • D.Placing the paper tray list inside the empty tray first
    Show answer

    Answer: C. Performing hand hygiene

    The WHO manual lists hand hygiene before inspection work and keeping the workbench clean and disinfected between sessions, to avoid contaminating clean devices.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  6. 6.During a function check of a toothed clamp, what should the technician confirm about the jaws?

    • A.The jaws are bent slightly apart so steam can get between them
    • B.The jaws have been coated with a light machine oil to prevent rust
    • C.The jaws stay locked on the last ratchet tooth while in storage
    • D.The jaws and teeth line up when closed
    Show answer

    Answer: D. The jaws and teeth line up when closed

    The WHO manual lists checking that jaws and teeth meet in line. Misaligned jaws do not grip properly and mean the instrument needs repair.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  7. 7.Which group of instruments should be checked for sharpness during inspection?

    • A.Hemostats, needle holders and towel clamps
    • B.Retractors, basins and vaginal specula
    • C.Suction tips, cannulas and trocar sleeves
    • D.Scissors, rongeurs, chisels and curettes
    Show answer

    Answer: D. Scissors, rongeurs, chisels and curettes

    The WHO manual lists cutting instruments such as scissors, rongeurs, chisels and curettes as items whose edges must be checked for sharpness.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  8. 8.Why are insulated laparoscopic and electrosurgical instruments tested for insulation failure?

    • A.Damaged insulation lets steam reach inner wiring and stops sterilization
    • B.The test confirms the inside of the shaft was thoroughly cleaned
    • C.Breaks in insulation can let current escape and burn the patient or staff
    • D.Torn insulation makes the instrument heavier and harder for the surgeon to grip
    Show answer

    Answer: C. Breaks in insulation can let current escape and burn the patient or staff

    Damaged insulation can let electrosurgical current leak to tissue or people, causing burns and fires. Many such burns go unnoticed during surgery, so instruments are tested before they are packaged.

    Checked against Healthcare Purchasing News, Inspecting electrical instruments for insulation failure

  9. 9.A technician notices a crack at the box lock of a needle holder. What is the correct action?

    • A.Remove it from the set, report it to the supervisor, and document it
    • B.Put it back in the set, placed at the bottom so it is the last one picked up
    • C.Throw it in the trash without telling anyone
    • D.Bend the joint back into shape with pliers before packaging
    Show answer

    Answer: A. Remove it from the set, report it to the supervisor, and document it

    The WHO manual says any damaged, incomplete or malfunctioning device should be reported to the supervisor immediately. It is removed from service and documented for repair.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  10. 10.Which type of lubricant is appropriate for surgical instruments that need it?

    • A.Silicone spray from the maintenance shop
    • B.Mineral oil or light machine oil
    • C.Water-soluble instrument lubricant
    • D.Petroleum jelly applied to the joints
    Show answer

    Answer: C. Water-soluble instrument lubricant

    The WHO manual says lubricants should be water-soluble and used according to the device manufacturer's instructions; oily substances should not be used.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  11. 11.A technician is unsure which parts of a powered instrument to lubricate and how much to use. Where should they look?

    • A.The device manufacturer's instructions for use
    • B.The lubricant brand the operating room prefers
    • C.Whatever amount the previous shift used
    • D.A general rule of one drop per joint for all devices
    Show answer

    Answer: A. The device manufacturer's instructions for use

    The WHO manual says the manufacturer's instructions govern which components need lubrication, which lubricant to use, and how to prepare and apply it, including amount and immersion time.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Ch. 5 and 8 (copy hosted by University of Washington)

  12. 12.What gives stainless steel instruments their resistance to rust?

    • A.A clear plastic coating applied at the factory
    • B.A thin passive film of chromium and iron oxides
    • C.A layer of lubricant baked on by repeated steam cycles
    • D.Nickel plating that is renewed during each wash
    Show answer

    Answer: B. A thin passive film of chromium and iron oxides

    A passive layer of chromium and iron oxides forms on stainless steel. When it is damaged and cannot re-form, the instrument becomes prone to corrosion.

    Checked against Infection Control Today, Spotting, Staining, and Corrosion of Surgical Instruments