Free CRCST Sterilization Process practice test
HSPA gives this domain 21% of the exam. Answer each question, read why every choice is right or wrong, and see your score by topic.
- Steam and High-Temperature Sterilization
- Low-Temperature Sterilization
- Sterilizer Testing, Load Release and Recalls
CRCST Sterilization Process
How does ethylene oxide kill microorganisms?
CRCST Sterilization Process practice questions and answers
Read each question, choose your answer, then open the explanation.
1.How does ethylene oxide kill microorganisms?
- A.By alkylating proteins, DNA and RNA
- B.By coagulating proteins with moist heat
- C.By oxidizing cell components with hot, dry air
- D.By forming free radicals with radio-frequency energy
Show answer
Answer: A. By alkylating proteins, DNA and RNA
CDC says ethylene oxide inactivates microorganisms by alkylating proteins, DNA and RNA, which blocks normal cell metabolism and replication. Bacterial spores are the most resistant to it.
Checked against CDC, Disinfection and Sterilization Guideline: Ethylene Oxide Sterilization
2.According to CDC's comparison table, which of these cannot be processed in a hydrogen peroxide gas plasma sterilizer?
- A.A dry stainless steel instrument in a polyolefin pouch
- B.A dry plastic device in a polypropylene wrap
- C.A container holding liquid
- D.A heat-sensitive electrical device that is fully dry
Show answer
Answer: C. A container holding liquid
CDC's table lists cellulose (paper), linens and liquids as items that cannot be processed by hydrogen peroxide gas plasma. The process needs dry loads and synthetic packaging such as polypropylene wraps and polyolefin pouches.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), PDF
3.What is OSHA's permissible exposure limit (PEL) for ethylene oxide?
- A.5 ppm as an 8-hour TWA
- B.1 ppm as an 8-hour TWA
- C.0.5 ppm over 15 minutes
- D.10 ppm as a ceiling value
Show answer
Answer: B. 1 ppm as an 8-hour TWA
Under 29 CFR 1910.1047, employee exposure must not exceed 1 ppm averaged over an 8-hour workday. Separate limits apply to short excursions.
Checked against OSHA, 29 CFR 1910.1047 Ethylene Oxide
4.Which description of ethylene oxide is accurate?
- A.A nonflammable gas whose only health effect is mild, short-term eye irritation
- B.A liquid sterilant that breaks down into harmless water vapor and oxygen
- C.A flammable gas that CDC says should be considered a known human carcinogen
- D.A gas that poses no hazard to staff once the sterilizer door is sealed
Show answer
Answer: C. A flammable gas that CDC says should be considered a known human carcinogen
EtO is a colorless, flammable and explosive gas. CDC states it should be considered a known human carcinogen, and it is linked to neurologic, reproductive and other harms.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Ethylene Oxide "Gas" Sterilization
5.Which packaging material cannot be used in a hydrogen peroxide gas plasma sterilizer?
- A.Polypropylene sterilization wrap
- B.Cellulose materials such as paper and linen
- C.Polyolefin (Tyvek) peel pouches
- D.The special container tray made for the system
Show answer
Answer: B. Cellulose materials such as paper and linen
Cellulose (paper), linens and liquids cannot be processed with hydrogen peroxide gas plasma because they absorb the peroxide. The method needs synthetic packaging such as polypropylene wraps and polyolefin pouches.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF
6.Why can items processed in a hydrogen peroxide gas plasma sterilizer be used right after the cycle?
- A.A built-in 8- to 12-hour aeration phase is already included in the cycle
- B.The byproducts are water vapor and oxygen, so no aeration is needed
- C.The items are rinsed with sterile water automatically at the end of the cycle
- D.Hydrogen peroxide residue is harmless on skin, so it may be left on items
Show answer
Answer: B. The byproducts are water vapor and oxygen, so no aeration is needed
CDC states the process byproducts, water vapor and oxygen, are nontoxic. No aeration is needed, so items can be handled right away for use or storage.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Hydrogen Peroxide Gas Plasma
7.Per CDC, what must happen to critical items processed by the liquid peracetic acid immersion system?
- A.They may be wrapped and stored up to 30 days
- B.They must be used immediately
- C.They go in a closed cabinet until needed
- D.They must aerate 12 hours before use
Show answer
Answer: B. They must be used immediately
CDC says critical items from the peracetic acid immersion system must go straight to use: they leave the processor unwrapped and are not fully protected from contamination, so storing them is unacceptable. It is a just-in-time, point-of-use system.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF
8.How does an ozone sterilizer get its sterilant?
- A.It draws it from a single-dose ethylene oxide gas cartridge
- B.It dilutes a 35% peracetic acid concentrate in a sealed cup
- C.It makes it inside the unit from oxygen, water and electricity
- D.It vaporizes hydrogen peroxide from a sealed cassette or cartridge
Show answer
Answer: C. It makes it inside the unit from oxygen, water and electricity
CDC states the ozone sterilizer makes its own sterilant inside the unit from USP-grade oxygen, water and electricity. When the cycle ends, a catalyst breaks the leftover ozone down into oxygen and water vapor before it is vented.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods
9.Per CDC, which is a common time-temperature setting for dry heat sterilization?
- A.170°C (340°F) for 60 minutes
- B.121°C (250°F) for 30 minutes
- C.132°C (270°F) for 4 minutes
- D.50°C (122°F) for 12 minutes
Show answer
Answer: A. 170°C (340°F) for 60 minutes
CDC lists 170°C for 60 minutes, 160°C for 120 minutes, or 150°C for 150 minutes as common dry heat settings. Without moisture, heat kills more slowly, so temperatures are higher and times longer.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods
10.What is OSHA's excursion limit for ethylene oxide?
- A.1 ppm averaged over 15 minutes
- B.0.5 ppm averaged over 8 hours
- C.50 ppm averaged over 1 hour
- D.5 ppm averaged over 15 minutes
Show answer
Answer: D. 5 ppm averaged over 15 minutes
OSHA sets an excursion limit of 5 ppm averaged over a 15-minute sampling period, in addition to the 1 ppm 8-hour TWA.
Checked against OSHA, 29 CFR 1910.1047 Ethylene Oxide
11.Which sterilization monitor is the only one that directly measures whether the process killed microorganisms?
- A.The internal chemical indicator
- B.The sterilizer printout of time, temperature and pressure
- C.The external indicator tape
- D.The biological indicator
Show answer
Answer: D. The biological indicator
CDC says biological indicators are the only monitors that measure the lethality of a sterilization process directly. They carry spores that resist the process better than the bioburden found on medical devices.
Checked against CDC, Disinfection and Sterilization Guideline: Sterilizing Practices
12.According to CDC, where should the biological indicator test pack be placed in an ethylene oxide sterilizer load?
- A.Flat on the front bottom shelf, near the drain
- B.On the top shelf, nearest the door
- C.In the center of the load
- D.Alone in an otherwise empty chamber
Show answer
Answer: C. In the center of the load
For ethylene oxide, CDC describes putting the BI in a plastic syringe, folding it into a clean surgical towel, wrapping it and placing it in the center of the load. The front-bottom spot near the drain is used for steam.
Checked against CDC, Disinfection and Sterilization Guideline: Sterilizing Practices
13.What should the operator do with the sterilizer printout when a cycle ends?
- A.File it unread with the day's records, since the sterilizer checks itself
- B.Look at it only if the external chemical indicators failed to change
- C.Throw it away once the next biological indicator result is negative
- D.Check that time, temperature and pressure were met, then initial it
Show answer
Answer: D. Check that time, temperature and pressure were met, then initial it
WHO says the printout is reviewed and initialed at cycle completion for critical elements such as exposure time, temperature and pressure, and the cycle number is checked against the load label.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Sterilization chapter
14.What does a chemical indicator on the outside of a package show?
- A.That the contents are sterile and safe to use
- B.That the package went through a sterilization process
- C.That the sterilant reached all the way to the center of the pack
- D.That the most resistant spores were killed
Show answer
Answer: B. That the package went through a sterilization process
CDC explains that external chemical indicators show a package has been through a sterilization cycle but do not prove sterilization was achieved. Internal indicators and BIs give more information.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
15.When and how is the daily Bowie-Dick test run?
- A.Once a week, in a full load with the routine BI
- B.After every load, on the top shelf with the packs
- C.Each day before the first load, in an empty chamber
- D.Once a month, with the preventive maintenance
Show answer
Answer: C. Each day before the first load, in an empty chamber
CDC says the Bowie-Dick test is run daily before the first processed load, in an otherwise empty chamber. A failed test means the sterilizer is not used until it is inspected and passes.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization
16.Which organism is used in biological indicators for steam sterilization?
- A.Geobacillus stearothermophilus
- B.Bacillus atrophaeus
- C.Clostridioides difficile
- D.Staphylococcus aureus
Show answer
Answer: A. Geobacillus stearothermophilus
CDC states G. stearothermophilus spores are used to monitor steam sterilization because they are highly resistant to moist heat.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
17.Which organism is used in biological indicators for ethylene oxide and dry heat?
- A.Geobacillus stearothermophilus
- B.Bacillus atrophaeus
- C.Mycobacterium terrae
- D.Pseudomonas aeruginosa
Show answer
Answer: B. Bacillus atrophaeus
CDC states B. atrophaeus spores are used to monitor EtO and dry heat sterilization.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
18.Per CDC, how often should a steam sterilizer be monitored with biological indicators at a minimum?
- A.Once a month, unless the chemical indicators fail
- B.Only after installation and after major repairs
- C.At least weekly, preferably daily when busy
- D.Only in loads that contain implantable devices
Show answer
Answer: C. At least weekly, preferably daily when busy
CDC says steam and low-temperature sterilizers should be monitored with BIs at least weekly. Daily use is advised for busy sterilizers so failures are found sooner and fewer items must be recalled.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
19.How should a load that contains an implantable device be monitored?
- A.With a BI in that load, holding the implant until the BI is negative when possible
- B.The routine weekly BI is enough to cover any implants processed that week
- C.No BI is needed as long as a Type 5 integrating indicator placed inside the set has passed
- D.Only the first implant load of each day needs a biological indicator
Show answer
Answer: A. With a BI in that load, holding the implant until the BI is negative when possible
CDC recommends a BI in every load containing implantable items and quarantining the implants, whenever possible, until the BI is negative.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF
20.Per CDC, what should the load label on a sterilized item identify?
- A.The sterilizer, the cycle or load number, and the sterilization date
- B.Only the expiration date, if one is used
- C.Only the initials of the technician
- D.Only the set name, the tray count and the number of instruments inside
Show answer
Answer: A. The sterilizer, the cycle or load number, and the sterilization date
CDC recommends a load label that names the sterilizer, the cycle or load number and the date it was sterilized, plus the expiration date when one is used. This lets the item be traced to its cycle for a recall.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF
21.According to CDC, which four parameters define a steam sterilization process?
- A.Steam, pressure, temperature and time
- B.Steam, relative humidity, gas concentration and time
- C.Temperature, time, load weight and chamber volume
- D.Pressure, vacuum depth, drying time and cooling time
Show answer
Answer: A. Steam, pressure, temperature and time
CDC lists the four parameters of steam sterilization as steam, pressure, temperature and time. Pressure lets the steam reach the high temperatures needed to kill microbes quickly, and time is how long the load is held there.
Checked against CDC, Disinfection and Sterilization Guideline: Steam Sterilization
22.An instrument comes out of an immediate-use steam cycle and must go straight to the sterile field. According to CDC, how can a burn to the patient be prevented?
- A.Rinse it under cool running tap water in the substerile room
- B.Wipe it down with an alcohol pad, which evaporates and cools it
- C.Let it air-cool, or immerse it in a sterile liquid such as saline
- D.Set it in a clean refrigerator for a few minutes before handing it off
Show answer
Answer: C. Let it air-cool, or immerse it in a sterile liquid such as saline
CDC reports patients burned by flash-sterilized instruments and says burns can be prevented by air-cooling the instruments or immersing them in a sterile liquid such as saline. Staff should carry hot trays with heat-protective gloves.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), PDF
23.How does a high-speed prevacuum steam sterilizer remove air from the chamber?
- A.A vacuum pump draws the air out before steam is let in
- B.Incoming steam pushes heavier air down and out the drain
- C.Steam pulses above atmospheric pressure flush the air out
- D.A fan circulates hot air until the steam displaces it
Show answer
Answer: A. A vacuum pump draws the air out before steam is let in
CDC describes prevacuum sterilizers as fitted with a vacuum pump that pulls air out of the chamber and the load before steam enters. Steam then penetrates almost instantly, even into porous loads.
Checked against CDC, Disinfection and Sterilization Guideline: Steam Sterilization
24.Per CDC, what is the minimum exposure time for wrapped healthcare supplies in a gravity-displacement steam sterilizer at 121°C (250°F)?
- A.4 minutes
- B.30 minutes
- C.15 minutes
- D.10 minutes
Show answer
Answer: B. 30 minutes
CDC lists 30 minutes at 121°C as the minimum exposure for wrapped supplies in a gravity-displacement sterilizer. Gravity air removal is slow, so the hold at the lower temperature is long.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization
25.Per CDC, what is the minimum exposure time for wrapped supplies in a prevacuum steam sterilizer at 132°C (270°F)?
- A.30 minutes
- B.15 minutes
- C.25 minutes
- D.4 minutes
Show answer
Answer: D. 4 minutes
CDC gives 4 minutes at 132°C as the minimum for wrapped supplies in a prevacuum sterilizer. The vacuum removes air before steam enters, so penetration is nearly immediate.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization
26.Why must air be removed from the chamber and from packages during a steam cycle?
- A.Air in the chamber reacts with steam and leaves a residue on instruments
- B.Trapped air keeps steam from directly contacting device surfaces
- C.Air raises chamber pressure above the setting of the safety relief valve
- D.Air cools the jacket enough to stop the door interlock from releasing
Show answer
Answer: B. Trapped air keeps steam from directly contacting device surfaces
Steam sterilizes only by direct contact with every surface, and that contact cannot happen while air remains in the devices, packs or chamber. Air pockets act as cold spots where organisms can survive.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Sterilization chapter
27.What is the job of the door gasket on a steam sterilizer?
- A.It strains lint and debris from condensate leaving the chamber
- B.It senses chamber temperature for the recorder or printout
- C.It pulls air out of packs during the conditioning phase
- D.It seals the door so steam does not leak out
Show answer
Answer: D. It seals the door so steam does not leak out
The gasket forms an airtight seal between the door and the chamber so steam and pressure stay inside. A damaged or dirty gasket can leak and should be serviced, then rechecked.
Checked against Tuttnauer Knowledge Center, Key Components of Autoclaves for Effective Sterilization
28.On a steam sterilizer, what is the jacket?
- A.A steam-filled space around the chamber that heats the chamber walls
- B.The rubber seal fitted into the groove around the door opening
- C.The screen over the chamber drain that catches lint and debris
- D.The recorder that prints time, temperature and pressure data
Show answer
Answer: A. A steam-filled space around the chamber that heats the chamber walls
Many steam sterilizers have a jacket, a space between the inner and outer walls filled with steam, that preheats the chamber. Warm walls reduce condensation inside the chamber.
Checked against Tuttnauer Knowledge Center, Key Components of Autoclaves for Effective Sterilization
29.How should peel pouches be positioned in a steam sterilizer load?
- A.Flat and stacked on top of each other with the paper side facing up
- B.On edge in a perforated or mesh-bottom basket or rack
- C.Flat on the bottom shelf, underneath the heavy instrument trays
- D.Inside a closed solid-bottom container so the seals are protected
Show answer
Answer: B. On edge in a perforated or mesh-bottom basket or rack
CDC's loading guidance is to stand peel packs up on their edges in baskets or racks with perforated or mesh bottoms. Upright pouches let steam circulate around them and let condensate drain away.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
30.How should basins and other nonperforated containers be placed in a steam sterilizer?
- A.Flat and upright on the top shelf
- B.Nested together with no separators
- C.Flat on the bottom shelf over the drain
- D.On their edge
Show answer
Answer: D. On their edge
CDC states that nonperforated containers such as basins should be placed on their edge. That position lets air escape and condensate drain out.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
31.After a steam cycle, a technician finds droplets of moisture inside a wrapped tray. What should happen to it?
- A.Consider it contaminated and reprocess it
- B.Let it air-dry on the cart, then shelve it
- C.Wipe the outside dry and send it to storage
- D.Release it if the internal indicator passed
Show answer
Answer: A. Consider it contaminated and reprocess it
CDC treats any sterile item that gets wet as contaminated, because moisture carries microbes from the air and surfaces through the packaging. Reprocess the pack and find out why it came out wet.
Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices
32.A steam load has just been removed from the sterilizer. Where should it cool?
- A.On the cart in a low-traffic area, away from vents, until it reaches room temperature
- B.On a stainless-steel counter so the metal pulls heat out of the packs quickly
- C.Under an air-conditioning vent so it reaches room temperature sooner
- D.Straight onto sterile storage shelves while still warm, to save time and extra handling
Show answer
Answer: A. On the cart in a low-traffic area, away from vents, until it reaches room temperature
WHO advises letting the load cool to room temperature on the cart in a quiet area with no foot traffic and no strong drafts of warm or cold air. How long that takes depends on what is in the load.
Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Sterilization chapter