This topic covers how steam sterilizers remove air, the phases of a cycle and the parts that run it, CDC's minimum exposure times, how to load and unload a cart, wet packs, and when immediate-use steam sterilization is allowed. Learn the reason behind each rule: steam must touch every surface, and a sterile item must stay dry and protected.
Why steam works and why air is the enemy
- CDC calls steam the most widely used and most dependable sterilization method; it is cheap, nontoxic and kills spores quickly.
- Moist heat kills by permanently coagulating and denaturing the proteins and enzymes microbes need.
- Steam is the preferred method for critical items that heat, moisture and pressure will not damage.
- Steam only sterilizes what it touches, so every surface of every item must be reached.
- Air left in the chamber, a pack or a device blocks that contact and slows heating, leaving cold spots where organisms can survive.
- Ideal steam is dry saturated steam with a dryness fraction of at least 97%; wetter steam carries liquid water into the load.
Three ways to remove air
- Gravity displacement: steam enters at the top or sides, and because it is lighter than air it pushes the air down and out a drain near the bottom.
- Gravity air removal is slow and incomplete in porous items, so gravity cycles need longer exposure times.
- Prevacuum (dynamic air removal): a vacuum pump or ejector pulls air out of the chamber and load before steam comes in, so steam penetrates porous loads almost at once.
- Steam-flush pressure-pulse: repeated steam flushes and pressure pulses drive air out while the chamber stays above atmospheric pressure.
- Because a pressure-pulse chamber is never under vacuum, small air leaks cannot pull room air in; a vacuum-type unit (pump or ejector) can be affected by leaks.
Cycle phases and sterilizer parts
- Conditioning comes first: air is removed and the load is heated.
- Exposure follows: the chamber holds the set temperature for the set time.
- Exhaust vents the steam, and drying removes moisture last.
- A prevacuum unit dries fabric loads faster by pulling one more vacuum at the end of the cycle; the vacuum pulses at the start are for air removal.
- Jacket: a steam-filled space around the chamber that preheats it and helps dry the load at the end of the cycle.
- Door gasket: the seal between door and chamber that keeps steam and pressure in; inspect and clean it daily, and report damage.
- Thermostatic trap: works with the chamber drain to let condensate out, which helps prevent wet packs.
- Drain strainer (screen): catches lint and debris; a clogged one keeps condensate from draining and can cause wet packs, so clean it daily per the sterilizer IFU before running loads.
- Pressure relief (safety) valve: protects the vessel and staff from over-pressure; it is not part of air removal or drying.
CDC minimum exposure times
These are minimums at a given temperature. The device, sterilizer and container or wrap manufacturers' instructions come first, and an IFU that calls for a longer or special cycle must be run exactly as written.
- Gravity, 121°C (250°F): 30 minutes for wrapped instruments, textile packs and wrapped utensils.
- Gravity, 132°C (270°F): 15 minutes for wrapped instruments and utensils, but 25 minutes for textile packs, because fabric holds air.
- Prevacuum, 132°C (270°F): 4 minutes for wrapped instruments, textile packs and wrapped utensils.
- Steam-flush pressure-pulse: CDC gives 3 to 4 minutes at 132 to 135°C for porous loads and instruments.
- Two short cycles run back to back do not equal one validated extended cycle.
Loading the cart
- Load loosely so steam can circulate around every item and condensate can drain.
- Perforated trays lie flat, parallel to the shelf.
- Basins and other nonperforated containers stand on edge so air can escape and water can run out.
- Tilt concave items so water does not pool in them.
- Peel pouches stand on edge in a perforated or mesh-bottom basket or rack.
- Only the paper face of a paper-plastic pouch lets steam in, so set each plastic face against the next pouch's paper face.
- Put heavy metal sets on the lower shelves and textile packs above them, so condensate from the metal does not drip onto soft goods.
- Keep packs off the chamber walls; contact stains and damages them and blocks steam flow.
- Do not stack rigid containers unless the container maker has validated that stacked setup.
Wet packs
- A pack with dampness, droplets or puddles on or inside it is a wet pack.
- A wet pack is considered contaminated: moisture lets microbes travel through the packaging, so reprocess it and find the cause.
- Drying it afterward, wiping the outside or a passed indicator does not make a wet pack usable.
- Causes include heavy metal mass, dense wrapping, set design, overloading, a blocked drain and wet steam.
- If one heavy set is wet while the rest of the load is dry, look first at that set's weight and design.
- If wet packs show up across many loads, suspect steam quality or the steam traps; a dryness fraction of at least 97% is the goal.
Unloading and cooling
- Check the printout and external indicators, and inspect each item for moisture or damaged packaging.
- Leave the load on the cart to cool to room temperature in a quiet, low-traffic spot away from drafts of hot or cold air.
- Do not touch packs until they are cool: a hot pack acts like a wick and can pull moisture and microbes from hands through the wrap.
- Never set warm packs on cold surfaces such as metal counters or near air vents; condensation can form and contaminate the contents.
Immediate-use (flash) steam sterilization
- Items are steam sterilized unwrapped, in an open tray or a rigid container cleared for this use, for immediate use.
- Wrapped sterilization is preferred, because flashed items come out wet, hot and unprotected and cannot go to sterile storage.
- CDC allows it for items needed at once, such as a dropped instrument with no sterile spare, and for items that cannot be packaged and stored ahead.
- CDC says not to use it for convenience, to save time or instead of buying more instrument sets.
- Do not flash implants unless it truly cannot be avoided; then record the load ID, the patient's name or hospital ID and the BI result.
- CDC examples at 132°C: gravity, 3 minutes for nonporous metal items only and 10 minutes when porous items or lumens are included; prevacuum, 3 minutes nonporous and 4 minutes mixed.
Key terms
- Gravity displacement
- Air removal in which incoming steam, lighter than air, pushes air down and out through a drain near the bottom.
- Dynamic air removal
- Active air removal before exposure, either by vacuum pulses (prevacuum) or by steam flushes and pressure pulses.
- Prevacuum sterilizer
- Steam unit that uses a vacuum pump or ejector to pull air out of the chamber and load before exposure.
- Steam-flush pressure-pulse
- Air-removal method using repeated steam flushes and pulses above atmospheric pressure, so leaks cannot draw air in.
- Conditioning
- First cycle phase, when air is removed and the load is heated.
- Exposure time
- Period the load is held at the set sterilizing temperature.
- Jacket
- Steam-filled space around the chamber that preheats it and helps dry the load.
- Door gasket
- Seal between the door and chamber that keeps steam and pressure inside.
- Thermostatic trap
- Drain-line device that lets condensate leave the chamber, helping loads dry.
- Drain strainer
- Screen over the chamber drain that catches lint and debris; clean it daily.
- Dryness fraction
- Share of steam that is vapor rather than liquid water; at least 97% is the goal.
- Wet pack
- Package with dampness, droplets or puddles on or inside it after the cycle; treated as contaminated.
- Immediate-use steam sterilization
- Unwrapped steam processing of an item needed right away; also called flash; no storage afterward.
- Porous item
- Item that can trap air or liquid, such as fabric, rubber, plastic or a device with a lumen.
Common mistakes
- Mixing up the time tables: 30 minutes at 121°C gravity, 15 minutes at 132°C gravity for instruments, 25 minutes at 132°C gravity for textiles, and 4 minutes at 132°C prevacuum. The 3 and 10 minute figures are flash examples, not wrapped cycles.
- Thinking steam dissolves or reacts with air. Air must be physically pushed or pumped out; trapped air simply blocks steam contact.
- Releasing or air-drying a wet pack because its internal indicator passed. A wet pack is contaminated and is reprocessed.
- Loading pouches flat or stacked, basins upright, packs against the wall, or heavy metal above textiles. Pouches and basins go on edge, perforated trays flat, heavy items low.
- Cooling warm packs fast on a metal counter or under a vent, or shelving them warm. They cool on the cart in a quiet area and are handled only when cool.
- Using flash sterilization to speed turnover, cover a late loaner or avoid buying sets, or flashing implants routinely because an integrator passed.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Flash Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), full PDF (recommendations, glossary, Tables 7 and 8); WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016); Tuttnauer, How Does an Autoclave Work; Tuttnauer, Key Components of Autoclaves for Effective Sterilization; Tuttnauer, Autoclave Troubleshooting: Sterilization Failures; Steelco Belimed, Medical Sterilizers: Routine Care and Maintenance