CRCST Steam and High-Temperature Sterilization: study notes and practice questions

Part of the CRCST Sterilization Process domain, 21% of the exam. This topic covers how steam sterilizers remove air, the phases of a cycle and the parts that run it, CDC's minimum exposure times, how to load and unload a cart, wet packs, and when immediate-use steam sterilization is allowed. Learn the reason behind each rule: steam must touch every surface, and a sterile item must stay dry and protected.

Study notes ↓Chapter quiz ↓All free questions ↓

All chapters

CRCST Steam and High-Temperature Sterilization study notes

This topic covers how steam sterilizers remove air, the phases of a cycle and the parts that run it, CDC's minimum exposure times, how to load and unload a cart, wet packs, and when immediate-use steam sterilization is allowed. Learn the reason behind each rule: steam must touch every surface, and a sterile item must stay dry and protected.

Why steam works and why air is the enemy

  • CDC calls steam the most widely used and most dependable sterilization method; it is cheap, nontoxic and kills spores quickly.
  • Moist heat kills by permanently coagulating and denaturing the proteins and enzymes microbes need.
  • Steam is the preferred method for critical items that heat, moisture and pressure will not damage.
  • Steam only sterilizes what it touches, so every surface of every item must be reached.
  • Air left in the chamber, a pack or a device blocks that contact and slows heating, leaving cold spots where organisms can survive.
  • Ideal steam is dry saturated steam with a dryness fraction of at least 97%; wetter steam carries liquid water into the load.

Three ways to remove air

  • Gravity displacement: steam enters at the top or sides, and because it is lighter than air it pushes the air down and out a drain near the bottom.
  • Gravity air removal is slow and incomplete in porous items, so gravity cycles need longer exposure times.
  • Prevacuum (dynamic air removal): a vacuum pump or ejector pulls air out of the chamber and load before steam comes in, so steam penetrates porous loads almost at once.
  • Steam-flush pressure-pulse: repeated steam flushes and pressure pulses drive air out while the chamber stays above atmospheric pressure.
  • Because a pressure-pulse chamber is never under vacuum, small air leaks cannot pull room air in; a vacuum-type unit (pump or ejector) can be affected by leaks.

Cycle phases and sterilizer parts

  • Conditioning comes first: air is removed and the load is heated.
  • Exposure follows: the chamber holds the set temperature for the set time.
  • Exhaust vents the steam, and drying removes moisture last.
  • A prevacuum unit dries fabric loads faster by pulling one more vacuum at the end of the cycle; the vacuum pulses at the start are for air removal.
  • Jacket: a steam-filled space around the chamber that preheats it and helps dry the load at the end of the cycle.
  • Door gasket: the seal between door and chamber that keeps steam and pressure in; inspect and clean it daily, and report damage.
  • Thermostatic trap: works with the chamber drain to let condensate out, which helps prevent wet packs.
  • Drain strainer (screen): catches lint and debris; a clogged one keeps condensate from draining and can cause wet packs, so clean it daily per the sterilizer IFU before running loads.
  • Pressure relief (safety) valve: protects the vessel and staff from over-pressure; it is not part of air removal or drying.

CDC minimum exposure times

These are minimums at a given temperature. The device, sterilizer and container or wrap manufacturers' instructions come first, and an IFU that calls for a longer or special cycle must be run exactly as written.

  • Gravity, 121°C (250°F): 30 minutes for wrapped instruments, textile packs and wrapped utensils.
  • Gravity, 132°C (270°F): 15 minutes for wrapped instruments and utensils, but 25 minutes for textile packs, because fabric holds air.
  • Prevacuum, 132°C (270°F): 4 minutes for wrapped instruments, textile packs and wrapped utensils.
  • Steam-flush pressure-pulse: CDC gives 3 to 4 minutes at 132 to 135°C for porous loads and instruments.
  • Two short cycles run back to back do not equal one validated extended cycle.

Loading the cart

  • Load loosely so steam can circulate around every item and condensate can drain.
  • Perforated trays lie flat, parallel to the shelf.
  • Basins and other nonperforated containers stand on edge so air can escape and water can run out.
  • Tilt concave items so water does not pool in them.
  • Peel pouches stand on edge in a perforated or mesh-bottom basket or rack.
  • Only the paper face of a paper-plastic pouch lets steam in, so set each plastic face against the next pouch's paper face.
  • Put heavy metal sets on the lower shelves and textile packs above them, so condensate from the metal does not drip onto soft goods.
  • Keep packs off the chamber walls; contact stains and damages them and blocks steam flow.
  • Do not stack rigid containers unless the container maker has validated that stacked setup.

Wet packs

  • A pack with dampness, droplets or puddles on or inside it is a wet pack.
  • A wet pack is considered contaminated: moisture lets microbes travel through the packaging, so reprocess it and find the cause.
  • Drying it afterward, wiping the outside or a passed indicator does not make a wet pack usable.
  • Causes include heavy metal mass, dense wrapping, set design, overloading, a blocked drain and wet steam.
  • If one heavy set is wet while the rest of the load is dry, look first at that set's weight and design.
  • If wet packs show up across many loads, suspect steam quality or the steam traps; a dryness fraction of at least 97% is the goal.

Unloading and cooling

  • Check the printout and external indicators, and inspect each item for moisture or damaged packaging.
  • Leave the load on the cart to cool to room temperature in a quiet, low-traffic spot away from drafts of hot or cold air.
  • Do not touch packs until they are cool: a hot pack acts like a wick and can pull moisture and microbes from hands through the wrap.
  • Never set warm packs on cold surfaces such as metal counters or near air vents; condensation can form and contaminate the contents.

Immediate-use (flash) steam sterilization

  • Items are steam sterilized unwrapped, in an open tray or a rigid container cleared for this use, for immediate use.
  • Wrapped sterilization is preferred, because flashed items come out wet, hot and unprotected and cannot go to sterile storage.
  • CDC allows it for items needed at once, such as a dropped instrument with no sterile spare, and for items that cannot be packaged and stored ahead.
  • CDC says not to use it for convenience, to save time or instead of buying more instrument sets.
  • Do not flash implants unless it truly cannot be avoided; then record the load ID, the patient's name or hospital ID and the BI result.
  • CDC examples at 132°C: gravity, 3 minutes for nonporous metal items only and 10 minutes when porous items or lumens are included; prevacuum, 3 minutes nonporous and 4 minutes mixed.

Key terms

Gravity displacement
Air removal in which incoming steam, lighter than air, pushes air down and out through a drain near the bottom.
Dynamic air removal
Active air removal before exposure, either by vacuum pulses (prevacuum) or by steam flushes and pressure pulses.
Prevacuum sterilizer
Steam unit that uses a vacuum pump or ejector to pull air out of the chamber and load before exposure.
Steam-flush pressure-pulse
Air-removal method using repeated steam flushes and pulses above atmospheric pressure, so leaks cannot draw air in.
Conditioning
First cycle phase, when air is removed and the load is heated.
Exposure time
Period the load is held at the set sterilizing temperature.
Jacket
Steam-filled space around the chamber that preheats it and helps dry the load.
Door gasket
Seal between the door and chamber that keeps steam and pressure inside.
Thermostatic trap
Drain-line device that lets condensate leave the chamber, helping loads dry.
Drain strainer
Screen over the chamber drain that catches lint and debris; clean it daily.
Dryness fraction
Share of steam that is vapor rather than liquid water; at least 97% is the goal.
Wet pack
Package with dampness, droplets or puddles on or inside it after the cycle; treated as contaminated.
Immediate-use steam sterilization
Unwrapped steam processing of an item needed right away; also called flash; no storage afterward.
Porous item
Item that can trap air or liquid, such as fabric, rubber, plastic or a device with a lumen.

Common mistakes

  • Mixing up the time tables: 30 minutes at 121°C gravity, 15 minutes at 132°C gravity for instruments, 25 minutes at 132°C gravity for textiles, and 4 minutes at 132°C prevacuum. The 3 and 10 minute figures are flash examples, not wrapped cycles.
  • Thinking steam dissolves or reacts with air. Air must be physically pushed or pumped out; trapped air simply blocks steam contact.
  • Releasing or air-drying a wet pack because its internal indicator passed. A wet pack is contaminated and is reprocessed.
  • Loading pouches flat or stacked, basins upright, packs against the wall, or heavy metal above textiles. Pouches and basins go on edge, perforated trays flat, heavy items low.
  • Cooling warm packs fast on a metal counter or under a vent, or shelving them warm. They cool on the cart in a quiet area and are handled only when cool.
  • Using flash sterilization to speed turnover, cover a late loaner or avoid buying sets, or flashing implants routinely because an integrator passed.

Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Flash Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), full PDF (recommendations, glossary, Tables 7 and 8); WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016); Tuttnauer, How Does an Autoclave Work; Tuttnauer, Key Components of Autoclaves for Effective Sterilization; Tuttnauer, Autoclave Troubleshooting: Sterilization Failures; Steelco Belimed, Medical Sterilizers: Routine Care and Maintenance

Chapter quiz

6 questions from this chapter, with an explanation after each answer.

1/6

Steam and High-Temperature Sterilization

According to CDC, which four parameters define a steam sterilization process?

Steam and High-Temperature Sterilization questions and answers

Read each question, choose your answer, then open the explanation.

  1. 1.According to CDC, which four parameters define a steam sterilization process?

    • A.Steam, pressure, temperature and time
    • B.Steam, relative humidity, gas concentration and time
    • C.Temperature, time, load weight and chamber volume
    • D.Pressure, vacuum depth, drying time and cooling time
    Show answer

    Answer: A. Steam, pressure, temperature and time

    CDC lists the four parameters of steam sterilization as steam, pressure, temperature and time. Pressure lets the steam reach the high temperatures needed to kill microbes quickly, and time is how long the load is held there.

    Checked against CDC, Disinfection and Sterilization Guideline: Steam Sterilization

  2. 2.An instrument comes out of an immediate-use steam cycle and must go straight to the sterile field. According to CDC, how can a burn to the patient be prevented?

    • A.Rinse it under cool running tap water in the substerile room
    • B.Wipe it down with an alcohol pad, which evaporates and cools it
    • C.Let it air-cool, or immerse it in a sterile liquid such as saline
    • D.Set it in a clean refrigerator for a few minutes before handing it off
    Show answer

    Answer: C. Let it air-cool, or immerse it in a sterile liquid such as saline

    CDC reports patients burned by flash-sterilized instruments and says burns can be prevented by air-cooling the instruments or immersing them in a sterile liquid such as saline. Staff should carry hot trays with heat-protective gloves.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), PDF

  3. 3.How does a high-speed prevacuum steam sterilizer remove air from the chamber?

    • A.A vacuum pump draws the air out before steam is let in
    • B.Incoming steam pushes heavier air down and out the drain
    • C.Steam pulses above atmospheric pressure flush the air out
    • D.A fan circulates hot air until the steam displaces it
    Show answer

    Answer: A. A vacuum pump draws the air out before steam is let in

    CDC describes prevacuum sterilizers as fitted with a vacuum pump that pulls air out of the chamber and the load before steam enters. Steam then penetrates almost instantly, even into porous loads.

    Checked against CDC, Disinfection and Sterilization Guideline: Steam Sterilization

  4. 4.Per CDC, what is the minimum exposure time for wrapped healthcare supplies in a gravity-displacement steam sterilizer at 121°C (250°F)?

    • A.4 minutes
    • B.30 minutes
    • C.15 minutes
    • D.10 minutes
    Show answer

    Answer: B. 30 minutes

    CDC lists 30 minutes at 121°C as the minimum exposure for wrapped supplies in a gravity-displacement sterilizer. Gravity air removal is slow, so the hold at the lower temperature is long.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization

  5. 5.Per CDC, what is the minimum exposure time for wrapped supplies in a prevacuum steam sterilizer at 132°C (270°F)?

    • A.30 minutes
    • B.15 minutes
    • C.25 minutes
    • D.4 minutes
    Show answer

    Answer: D. 4 minutes

    CDC gives 4 minutes at 132°C as the minimum for wrapped supplies in a prevacuum sterilizer. The vacuum removes air before steam enters, so penetration is nearly immediate.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Steam Sterilization

  6. 6.Why must air be removed from the chamber and from packages during a steam cycle?

    • A.Air in the chamber reacts with steam and leaves a residue on instruments
    • B.Trapped air keeps steam from directly contacting device surfaces
    • C.Air raises chamber pressure above the setting of the safety relief valve
    • D.Air cools the jacket enough to stop the door interlock from releasing
    Show answer

    Answer: B. Trapped air keeps steam from directly contacting device surfaces

    Steam sterilizes only by direct contact with every surface, and that contact cannot happen while air remains in the devices, packs or chamber. Air pockets act as cold spots where organisms can survive.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Sterilization chapter

  7. 7.What is the job of the door gasket on a steam sterilizer?

    • A.It strains lint and debris from condensate leaving the chamber
    • B.It senses chamber temperature for the recorder or printout
    • C.It pulls air out of packs during the conditioning phase
    • D.It seals the door so steam does not leak out
    Show answer

    Answer: D. It seals the door so steam does not leak out

    The gasket forms an airtight seal between the door and the chamber so steam and pressure stay inside. A damaged or dirty gasket can leak and should be serviced, then rechecked.

    Checked against Tuttnauer Knowledge Center, Key Components of Autoclaves for Effective Sterilization

  8. 8.On a steam sterilizer, what is the jacket?

    • A.A steam-filled space around the chamber that heats the chamber walls
    • B.The rubber seal fitted into the groove around the door opening
    • C.The screen over the chamber drain that catches lint and debris
    • D.The recorder that prints time, temperature and pressure data
    Show answer

    Answer: A. A steam-filled space around the chamber that heats the chamber walls

    Many steam sterilizers have a jacket, a space between the inner and outer walls filled with steam, that preheats the chamber. Warm walls reduce condensation inside the chamber.

    Checked against Tuttnauer Knowledge Center, Key Components of Autoclaves for Effective Sterilization

  9. 9.How should peel pouches be positioned in a steam sterilizer load?

    • A.Flat and stacked on top of each other with the paper side facing up
    • B.On edge in a perforated or mesh-bottom basket or rack
    • C.Flat on the bottom shelf, underneath the heavy instrument trays
    • D.Inside a closed solid-bottom container so the seals are protected
    Show answer

    Answer: B. On edge in a perforated or mesh-bottom basket or rack

    CDC's loading guidance is to stand peel packs up on their edges in baskets or racks with perforated or mesh bottoms. Upright pouches let steam circulate around them and let condensate drain away.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  10. 10.How should basins and other nonperforated containers be placed in a steam sterilizer?

    • A.Flat and upright on the top shelf
    • B.Nested together with no separators
    • C.Flat on the bottom shelf over the drain
    • D.On their edge
    Show answer

    Answer: D. On their edge

    CDC states that nonperforated containers such as basins should be placed on their edge. That position lets air escape and condensate drain out.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  11. 11.After a steam cycle, a technician finds droplets of moisture inside a wrapped tray. What should happen to it?

    • A.Consider it contaminated and reprocess it
    • B.Let it air-dry on the cart, then shelve it
    • C.Wipe the outside dry and send it to storage
    • D.Release it if the internal indicator passed
    Show answer

    Answer: A. Consider it contaminated and reprocess it

    CDC treats any sterile item that gets wet as contaminated, because moisture carries microbes from the air and surfaces through the packaging. Reprocess the pack and find out why it came out wet.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilizing Practices

  12. 12.A steam load has just been removed from the sterilizer. Where should it cool?

    • A.On the cart in a low-traffic area, away from vents, until it reaches room temperature
    • B.On a stainless-steel counter so the metal pulls heat out of the packs quickly
    • C.Under an air-conditioning vent so it reaches room temperature sooner
    • D.Straight onto sterile storage shelves while still warm, to save time and extra handling
    Show answer

    Answer: A. On the cart in a low-traffic area, away from vents, until it reaches room temperature

    WHO advises letting the load cool to room temperature on the cart in a quiet area with no foot traffic and no strong drafts of warm or cold air. How long that takes depends on what is in the load.

    Checked against WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016), Sterilization chapter