This topic covers the methods used for items that heat or moisture would damage: ethylene oxide, hydrogen peroxide gas plasma, liquid peracetic acid and ozone, plus dry heat for items steam cannot reach. For each method, know what it can and cannot process, what packaging it needs, what happens after the cycle, and the safety rules that protect staff and patients.
Choosing the method
- Heat-resistant critical items go to steam, which has the widest safety margin.
- Critical items that are heat- or moisture-sensitive need a low-temperature method such as ethylene oxide (EtO), hydrogen peroxide gas plasma or peracetic acid.
- Compatibility differs for every method, so get written compatibility information from the device maker and efficacy data from the sterilizer maker.
- If a device IFU validates only one method, use that method; general compatibility claims for a sterilizer do not replace the device maker's validation.
- Running an unvalidated process twice does not validate it, and a heat-sensitive device must not be put through steam or flash instead.
- Narrow lumens challenge every low-temperature method, because the sterilant must directly touch the contaminated surface.
Ethylene oxide: the process
- EtO is a colorless gas that can burn and explode, and CDC says to treat it as a known human carcinogen.
- Four variables control it: how much gas, how hot, how humid, and for how long.
- CDC's ranges: about 37 to 63°C, 40 to 80% relative humidity and 1 to 6 hours of exposure.
- Cycle stages in order: preconditioning and humidification, gas introduction, exposure, evacuation, then air washes.
- Its strength is that it penetrates packaging and lumens and does not harm heat- and moisture-sensitive materials; its drawbacks are long cycles, cost and hazards.
Ethylene oxide: aeration
- Many materials absorb EtO, so items must be aerated before use to let the gas leave them.
- EtO residues left in implants have caused tissue burns in patients.
- In a mechanical aerator, plan on 8-12 hours at 50-60°C; at room temperature (20°C) it takes about 7 days.
- CDC's example: polyvinyl chloride (PVC) tubing needs 12 hours at 50°C or 8 hours at 60°C.
- Follow the device maker's aeration instructions.
- Never interrupt an aeration cycle or open the chamber to pull out a device, even when a surgeon asks; tell the team it is not available until aeration ends.
- A chemical indicator shows the item was exposed to the cycle, not that aeration is complete, and rinsing does not remove absorbed gas.
Ethylene oxide: worker safety (OSHA 29 CFR 1910.1047)
- Permissible exposure limit (PEL): 1 ppm, averaged over an 8-hour shift (8-hour TWA).
- Excursion limit: 5 ppm as a 15-minute average.
- Action level: 0.5 ppm as an 8-hour TWA.
- Workers whose exposure reaches or exceeds the action level on 30 or more days a year must get medical surveillance, by or under a licensed physician, at no cost to them.
- Workers exposed in an emergency release must also be offered medical exams.
- CDC notes that EtO cartridges should be kept in a flammable liquid storage cabinet.
Hydrogen peroxide gas plasma
- The chamber is pulled to a deep vacuum, peroxide is vaporized into it, and radio-frequency energy then creates the plasma.
- It runs below 50°C, with cycles of about 28 to 75 minutes depending on the model.
- Byproducts are water vapor and oxygen, which are nontoxic, so no aeration is needed and items can be used or stored right away.
- Items must be completely dry: moisture can keep the chamber from reaching vacuum, and the cycle aborts.
- Paper and other cellulose, linens and liquids cannot be processed because they absorb the peroxide.
- Use synthetic packaging validated for the method, such as polypropylene wrap, polyolefin (Tyvek) pouches and the system's special tray; never a paper-plastic steam pouch.
- Some long or narrow lumens cannot be processed; check that both the device IFU and the sterilizer maker's lumen length and diameter limits cover the device.
- Handle, store and dispose of sterilant cassettes as the sterilant and sterilizer makers direct; hydrogen peroxide may be toxic above 1 ppm TWA.
Liquid peracetic acid
- An automated system dilutes a single-dose 35% peracetic acid concentrate to 0.2% at about 50°C.
- It is for immersible, heat-sensitive instruments such as endoscopes; non-immersible electronics and textiles do not belong in it.
- Lumened scopes must be attached with the correct channel connectors so sterilant flows through every channel; infections followed use of the wrong connectors.
- The used sterilant is discarded, and the instruments are rinsed four times with filtered water after the cycle.
- It is a point-of-use system: items are not protected after the cycle, so they must be used immediately and never stored.
Ozone and dry heat
- Ozone sterilizers make their sterilant inside the unit using electricity, USP-grade oxygen and water of steam quality.
- At the end of an ozone cycle, a catalyst turns the ozone back into oxygen and water vapor.
- Dry heat is for materials moist heat would damage or cannot penetrate, such as powders, petroleum products and oils, and sharp instruments.
- Without moisture heat kills more slowly, so CDC's settings are hotter and longer than steam: 60 minutes at 170°C (340°F), 120 minutes at 160°C, or 150 minutes at 150°C.
- CDC notes those high temperatures are unsuitable for most materials, so dry heat is never an option for heat-sensitive devices.
Key terms
- Ethylene oxide (EtO)
- Flammable, carcinogenic sterilant gas used for heat- and moisture-sensitive devices; requires aeration afterward.
- Aeration
- Holding items after an EtO cycle, in an aerator or room, until the gas they absorbed has left them.
- Preconditioning
- First EtO cycle stage, when the load is brought to set conditions and humidified before gas enters.
- Permissible exposure limit (PEL)
- OSHA's maximum allowed worker exposure; for EtO, 1 ppm averaged over 8 hours.
- Excursion limit
- OSHA's short-term EtO cap: 5 ppm averaged over 15 minutes.
- Action level
- EtO exposure of 0.5 ppm over 8 hours; reaching it 30 or more days a year requires medical surveillance.
- Time-weighted average (TWA)
- A worker's average airborne chemical concentration across a set sampling period, such as an 8-hour shift.
- Hydrogen peroxide gas plasma
- Low-temperature vacuum process using peroxide vapor and plasma; leaves water and oxygen, so no aeration.
- Cellulose
- Plant-based material such as paper or cotton linen; absorbs hydrogen peroxide, so it cannot go in peroxide sterilizers.
- Polyolefin pouch
- Synthetic (Tyvek-type) peel pouch acceptable for hydrogen peroxide sterilization.
- Peracetic acid system
- Automated liquid immersion sterilizer for immersible heat-sensitive instruments; items must be used immediately.
- Channel connector
- Fitting that links an endoscope's lumens to the processor so liquid sterilant flows through them.
- Ozone sterilizer
- Low-temperature unit that generates its oxidizing gas from oxygen, water and electricity, then converts it back with a catalyst.
- Dry heat
- Hot-air sterilization, e.g. 170°C for 60 minutes, for powders, oils and petroleum products.
Common mistakes
- Swapping the EtO numbers: 1 ppm is the 8-hour PEL, 5 ppm over 15 minutes is the excursion limit, and 0.5 ppm is the action level. Medical surveillance starts at the action level for 30 or more days a year, not after one excursion.
- Pulling an item out of the aerator early because the indicator changed or 'most of the gas is gone'. Aeration runs its full time; PVC tubing needs 12 hours at 50°C or 8 hours at 60°C.
- Putting paper, linen, liquids or wet items in a hydrogen peroxide sterilizer, or assuming a longer cycle overcomes a lumen that is outside the stated limits.
- Thinking hydrogen peroxide loads need aeration (that is EtO) or a rinse (that is peracetic acid). Peroxide plasma leaves only water vapor and oxygen.
- Storing items from the liquid peracetic acid system or skipping the channel connectors. Items go straight to use, and every channel must be connected.
- Substituting a method the device IFU does not list, or sending a heat-sensitive device to steam, flash or dry heat because it is faster.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Ethylene Oxide Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Hydrogen Peroxide Gas Plasma; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Peracetic Acid Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), full PDF (recommendations and Table 6); OSHA, 29 CFR 1910.1047 Ethylene Oxide; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016)