CRCST Low-Temperature Sterilization: study notes and practice questions

Part of the CRCST Sterilization Process domain, 21% of the exam. This topic covers the methods used for items that heat or moisture would damage: ethylene oxide, hydrogen peroxide gas plasma, liquid peracetic acid and ozone, plus dry heat for items steam cannot reach. For each method, know what it can and cannot process, what packaging it needs, what happens after the cycle, and the safety rules that protect staff and patients.

Study notes ↓Chapter quiz ↓All free questions ↓

All chapters

CRCST Low-Temperature Sterilization study notes

This topic covers the methods used for items that heat or moisture would damage: ethylene oxide, hydrogen peroxide gas plasma, liquid peracetic acid and ozone, plus dry heat for items steam cannot reach. For each method, know what it can and cannot process, what packaging it needs, what happens after the cycle, and the safety rules that protect staff and patients.

Choosing the method

  • Heat-resistant critical items go to steam, which has the widest safety margin.
  • Critical items that are heat- or moisture-sensitive need a low-temperature method such as ethylene oxide (EtO), hydrogen peroxide gas plasma or peracetic acid.
  • Compatibility differs for every method, so get written compatibility information from the device maker and efficacy data from the sterilizer maker.
  • If a device IFU validates only one method, use that method; general compatibility claims for a sterilizer do not replace the device maker's validation.
  • Running an unvalidated process twice does not validate it, and a heat-sensitive device must not be put through steam or flash instead.
  • Narrow lumens challenge every low-temperature method, because the sterilant must directly touch the contaminated surface.

Ethylene oxide: the process

  • EtO is a colorless gas that can burn and explode, and CDC says to treat it as a known human carcinogen.
  • Four variables control it: how much gas, how hot, how humid, and for how long.
  • CDC's ranges: about 37 to 63°C, 40 to 80% relative humidity and 1 to 6 hours of exposure.
  • Cycle stages in order: preconditioning and humidification, gas introduction, exposure, evacuation, then air washes.
  • Its strength is that it penetrates packaging and lumens and does not harm heat- and moisture-sensitive materials; its drawbacks are long cycles, cost and hazards.

Ethylene oxide: aeration

  • Many materials absorb EtO, so items must be aerated before use to let the gas leave them.
  • EtO residues left in implants have caused tissue burns in patients.
  • In a mechanical aerator, plan on 8-12 hours at 50-60°C; at room temperature (20°C) it takes about 7 days.
  • CDC's example: polyvinyl chloride (PVC) tubing needs 12 hours at 50°C or 8 hours at 60°C.
  • Follow the device maker's aeration instructions.
  • Never interrupt an aeration cycle or open the chamber to pull out a device, even when a surgeon asks; tell the team it is not available until aeration ends.
  • A chemical indicator shows the item was exposed to the cycle, not that aeration is complete, and rinsing does not remove absorbed gas.

Ethylene oxide: worker safety (OSHA 29 CFR 1910.1047)

  • Permissible exposure limit (PEL): 1 ppm, averaged over an 8-hour shift (8-hour TWA).
  • Excursion limit: 5 ppm as a 15-minute average.
  • Action level: 0.5 ppm as an 8-hour TWA.
  • Workers whose exposure reaches or exceeds the action level on 30 or more days a year must get medical surveillance, by or under a licensed physician, at no cost to them.
  • Workers exposed in an emergency release must also be offered medical exams.
  • CDC notes that EtO cartridges should be kept in a flammable liquid storage cabinet.

Hydrogen peroxide gas plasma

  • The chamber is pulled to a deep vacuum, peroxide is vaporized into it, and radio-frequency energy then creates the plasma.
  • It runs below 50°C, with cycles of about 28 to 75 minutes depending on the model.
  • Byproducts are water vapor and oxygen, which are nontoxic, so no aeration is needed and items can be used or stored right away.
  • Items must be completely dry: moisture can keep the chamber from reaching vacuum, and the cycle aborts.
  • Paper and other cellulose, linens and liquids cannot be processed because they absorb the peroxide.
  • Use synthetic packaging validated for the method, such as polypropylene wrap, polyolefin (Tyvek) pouches and the system's special tray; never a paper-plastic steam pouch.
  • Some long or narrow lumens cannot be processed; check that both the device IFU and the sterilizer maker's lumen length and diameter limits cover the device.
  • Handle, store and dispose of sterilant cassettes as the sterilant and sterilizer makers direct; hydrogen peroxide may be toxic above 1 ppm TWA.

Liquid peracetic acid

  • An automated system dilutes a single-dose 35% peracetic acid concentrate to 0.2% at about 50°C.
  • It is for immersible, heat-sensitive instruments such as endoscopes; non-immersible electronics and textiles do not belong in it.
  • Lumened scopes must be attached with the correct channel connectors so sterilant flows through every channel; infections followed use of the wrong connectors.
  • The used sterilant is discarded, and the instruments are rinsed four times with filtered water after the cycle.
  • It is a point-of-use system: items are not protected after the cycle, so they must be used immediately and never stored.

Ozone and dry heat

  • Ozone sterilizers make their sterilant inside the unit using electricity, USP-grade oxygen and water of steam quality.
  • At the end of an ozone cycle, a catalyst turns the ozone back into oxygen and water vapor.
  • Dry heat is for materials moist heat would damage or cannot penetrate, such as powders, petroleum products and oils, and sharp instruments.
  • Without moisture heat kills more slowly, so CDC's settings are hotter and longer than steam: 60 minutes at 170°C (340°F), 120 minutes at 160°C, or 150 minutes at 150°C.
  • CDC notes those high temperatures are unsuitable for most materials, so dry heat is never an option for heat-sensitive devices.

Key terms

Ethylene oxide (EtO)
Flammable, carcinogenic sterilant gas used for heat- and moisture-sensitive devices; requires aeration afterward.
Aeration
Holding items after an EtO cycle, in an aerator or room, until the gas they absorbed has left them.
Preconditioning
First EtO cycle stage, when the load is brought to set conditions and humidified before gas enters.
Permissible exposure limit (PEL)
OSHA's maximum allowed worker exposure; for EtO, 1 ppm averaged over 8 hours.
Excursion limit
OSHA's short-term EtO cap: 5 ppm averaged over 15 minutes.
Action level
EtO exposure of 0.5 ppm over 8 hours; reaching it 30 or more days a year requires medical surveillance.
Time-weighted average (TWA)
A worker's average airborne chemical concentration across a set sampling period, such as an 8-hour shift.
Hydrogen peroxide gas plasma
Low-temperature vacuum process using peroxide vapor and plasma; leaves water and oxygen, so no aeration.
Cellulose
Plant-based material such as paper or cotton linen; absorbs hydrogen peroxide, so it cannot go in peroxide sterilizers.
Polyolefin pouch
Synthetic (Tyvek-type) peel pouch acceptable for hydrogen peroxide sterilization.
Peracetic acid system
Automated liquid immersion sterilizer for immersible heat-sensitive instruments; items must be used immediately.
Channel connector
Fitting that links an endoscope's lumens to the processor so liquid sterilant flows through them.
Ozone sterilizer
Low-temperature unit that generates its oxidizing gas from oxygen, water and electricity, then converts it back with a catalyst.
Dry heat
Hot-air sterilization, e.g. 170°C for 60 minutes, for powders, oils and petroleum products.

Common mistakes

  • Swapping the EtO numbers: 1 ppm is the 8-hour PEL, 5 ppm over 15 minutes is the excursion limit, and 0.5 ppm is the action level. Medical surveillance starts at the action level for 30 or more days a year, not after one excursion.
  • Pulling an item out of the aerator early because the indicator changed or 'most of the gas is gone'. Aeration runs its full time; PVC tubing needs 12 hours at 50°C or 8 hours at 60°C.
  • Putting paper, linen, liquids or wet items in a hydrogen peroxide sterilizer, or assuming a longer cycle overcomes a lumen that is outside the stated limits.
  • Thinking hydrogen peroxide loads need aeration (that is EtO) or a rinse (that is peracetic acid). Peroxide plasma leaves only water vapor and oxygen.
  • Storing items from the liquid peracetic acid system or skipping the channel connectors. Items go straight to use, and every channel must be connected.
  • Substituting a method the device IFU does not list, or sending a heat-sensitive device to steam, flash or dry heat because it is faster.

Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Ethylene Oxide Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Hydrogen Peroxide Gas Plasma; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Peracetic Acid Sterilization; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008), full PDF (recommendations and Table 6); OSHA, 29 CFR 1910.1047 Ethylene Oxide; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016)

Chapter quiz

6 questions from this chapter, with an explanation after each answer.

1/6

Low-Temperature Sterilization

How does ethylene oxide kill microorganisms?

Low-Temperature Sterilization questions and answers

Read each question, choose your answer, then open the explanation.

  1. 1.How does ethylene oxide kill microorganisms?

    • A.By alkylating proteins, DNA and RNA
    • B.By coagulating proteins with moist heat
    • C.By oxidizing cell components with hot, dry air
    • D.By forming free radicals with radio-frequency energy
    Show answer

    Answer: A. By alkylating proteins, DNA and RNA

    CDC says ethylene oxide inactivates microorganisms by alkylating proteins, DNA and RNA, which blocks normal cell metabolism and replication. Bacterial spores are the most resistant to it.

    Checked against CDC, Disinfection and Sterilization Guideline: Ethylene Oxide Sterilization

  2. 2.According to CDC's comparison table, which of these cannot be processed in a hydrogen peroxide gas plasma sterilizer?

    • A.A dry stainless steel instrument in a polyolefin pouch
    • B.A dry plastic device in a polypropylene wrap
    • C.A container holding liquid
    • D.A heat-sensitive electrical device that is fully dry
    Show answer

    Answer: C. A container holding liquid

    CDC's table lists cellulose (paper), linens and liquids as items that cannot be processed by hydrogen peroxide gas plasma. The process needs dry loads and synthetic packaging such as polypropylene wraps and polyolefin pouches.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated June 2024), PDF

  3. 3.What is OSHA's permissible exposure limit (PEL) for ethylene oxide?

    • A.5 ppm as an 8-hour TWA
    • B.1 ppm as an 8-hour TWA
    • C.0.5 ppm over 15 minutes
    • D.10 ppm as a ceiling value
    Show answer

    Answer: B. 1 ppm as an 8-hour TWA

    Under 29 CFR 1910.1047, employee exposure must not exceed 1 ppm averaged over an 8-hour workday. Separate limits apply to short excursions.

    Checked against OSHA, 29 CFR 1910.1047 Ethylene Oxide

  4. 4.Which description of ethylene oxide is accurate?

    • A.A nonflammable gas whose only health effect is mild, short-term eye irritation
    • B.A liquid sterilant that breaks down into harmless water vapor and oxygen
    • C.A flammable gas that CDC says should be considered a known human carcinogen
    • D.A gas that poses no hazard to staff once the sterilizer door is sealed
    Show answer

    Answer: C. A flammable gas that CDC says should be considered a known human carcinogen

    EtO is a colorless, flammable and explosive gas. CDC states it should be considered a known human carcinogen, and it is linked to neurologic, reproductive and other harms.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Ethylene Oxide "Gas" Sterilization

  5. 5.Which packaging material cannot be used in a hydrogen peroxide gas plasma sterilizer?

    • A.Polypropylene sterilization wrap
    • B.Cellulose materials such as paper and linen
    • C.Polyolefin (Tyvek) peel pouches
    • D.The special container tray made for the system
    Show answer

    Answer: B. Cellulose materials such as paper and linen

    Cellulose (paper), linens and liquids cannot be processed with hydrogen peroxide gas plasma because they absorb the peroxide. The method needs synthetic packaging such as polypropylene wraps and polyolefin pouches.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF

  6. 6.Why can items processed in a hydrogen peroxide gas plasma sterilizer be used right after the cycle?

    • A.A built-in 8- to 12-hour aeration phase is already included in the cycle
    • B.The byproducts are water vapor and oxygen, so no aeration is needed
    • C.The items are rinsed with sterile water automatically at the end of the cycle
    • D.Hydrogen peroxide residue is harmless on skin, so it may be left on items
    Show answer

    Answer: B. The byproducts are water vapor and oxygen, so no aeration is needed

    CDC states the process byproducts, water vapor and oxygen, are nontoxic. No aeration is needed, so items can be handled right away for use or storage.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Hydrogen Peroxide Gas Plasma

  7. 7.Per CDC, what must happen to critical items processed by the liquid peracetic acid immersion system?

    • A.They may be wrapped and stored up to 30 days
    • B.They must be used immediately
    • C.They go in a closed cabinet until needed
    • D.They must aerate 12 hours before use
    Show answer

    Answer: B. They must be used immediately

    CDC says critical items from the peracetic acid immersion system must go straight to use: they leave the processor unwrapped and are not fully protected from contamination, so storing them is unacceptable. It is a just-in-time, point-of-use system.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, last updated 2024), full PDF

  8. 8.How does an ozone sterilizer get its sterilant?

    • A.It draws it from a single-dose ethylene oxide gas cartridge
    • B.It dilutes a 35% peracetic acid concentrate in a sealed cup
    • C.It makes it inside the unit from oxygen, water and electricity
    • D.It vaporizes hydrogen peroxide from a sealed cassette or cartridge
    Show answer

    Answer: C. It makes it inside the unit from oxygen, water and electricity

    CDC states the ozone sterilizer makes its own sterilant inside the unit from USP-grade oxygen, water and electricity. When the cycle ends, a catalyst breaks the leftover ozone down into oxygen and water vapor before it is vented.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods

  9. 9.Per CDC, which is a common time-temperature setting for dry heat sterilization?

    • A.170°C (340°F) for 60 minutes
    • B.121°C (250°F) for 30 minutes
    • C.132°C (270°F) for 4 minutes
    • D.50°C (122°F) for 12 minutes
    Show answer

    Answer: A. 170°C (340°F) for 60 minutes

    CDC lists 170°C for 60 minutes, 160°C for 120 minutes, or 150°C for 150 minutes as common dry heat settings. Without moisture, heat kills more slowly, so temperatures are higher and times longer.

    Checked against CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008): Other Sterilization Methods

  10. 10.What is OSHA's excursion limit for ethylene oxide?

    • A.1 ppm averaged over 15 minutes
    • B.0.5 ppm averaged over 8 hours
    • C.50 ppm averaged over 1 hour
    • D.5 ppm averaged over 15 minutes
    Show answer

    Answer: D. 5 ppm averaged over 15 minutes

    OSHA sets an excursion limit of 5 ppm averaged over a 15-minute sampling period, in addition to the 1 ppm 8-hour TWA.

    Checked against OSHA, 29 CFR 1910.1047 Ethylene Oxide