This topic covers reusable patient care equipment such as infusion pumps, compression pumps, warming units, suction units and stretchers: how they are classified, cleaned and disinfected between patients, checked for electrical and maintenance problems, tracked, stored and sent for repair. Most answers come back to three rules: clean before you disinfect, follow the label and the IFU, and never issue equipment that is unsafe.
Classification and level of processing
- Under the Spaulding scheme, items that touch only intact skin, or do not touch the patient at all, are **noncritical**. Most patient care equipment, such as pumps, IV poles and blood pressure cuffs, falls here.
- Semicritical items touch mucous membranes or nonintact skin; critical items enter sterile tissue or the vascular system.
- Spaulding has no exempt class. Noncritical items still need cleaning and disinfection.
- CDC recommends low-level disinfection for noncritical equipment, after cleaning. High-level disinfection is the minimum for semicritical items, and sterilization is for critical items, so steam sterilizing a pump is not the answer.
Cleaning and disinfecting between patients
- CDC core practices: clean and reprocess reusable equipment before it is used on another patient, and whenever it is soiled. Monthly maintenance or contact-precaution status does not change this, and equipment can carry germs with no visible soil.
- Clean off visible soil first, then apply the disinfectant. Soil left in place can shield microbes from the chemical.
- Use an EPA-registered hospital disinfectant and follow its label. The surface must stay wet for the label's contact time, which varies by product. How fast the surface looks dry is not the test.
- Check the manufacturer's IFU before cleaning. CDC says manufacturers should state whether equipment is water-resistant or safe to immerse, and which germicides it tolerates. Never soak a pump just because a sink is deep enough or a coworker did it before.
- 70% isopropyl alcohol evaporates fast, so long contact times are hard to reach on large surfaces like a stretcher. It can also harm some rubber and plastics. It is not a high-level disinfectant.
- Remove and discard single-patient parts, such as compression sleeves, before cleaning the reusable pump. They are not cleaned and sent to the next patient.
Protecting yourself while handling returns
- OSHA's Bloodborne Pathogens standard requires gloves whenever a worker can reasonably expect hand contact with blood or other possibly infectious material, including when handling contaminated items or surfaces.
- Dried blood still counts. Add face and eye protection if splashes or sprays are likely.
One-way flow and clean storage
- Returned equipment goes to a soiled area kept apart from clean, ready-to-issue equipment. CDC core practices call for keeping clean and soiled equipment apart so contamination cannot pass between them.
- Equipment moves one way: soiled receiving, then decontamination, then clean storage, then out to the units. Contaminated items never travel back into clean areas.
- Soiled equipment never goes into sterile storage or straight back onto a distribution cart, and clean and soiled items do not share a cart.
- Store cleaned equipment in a clean, dry area away from soiled items and protected from moisture and dust. Not in the soiled room, not under a sink and not on a hallway floor.
- Charge rechargeable batteries as the IFU directs so units are ready when issued.
Electrical safety (OSHA 29 CFR 1910.334)
- Visually inspect cord-and-plug equipment before use for external defects: loose parts, bent or missing pins, cracks, exposed wires or a damaged cord jacket.
- This check happens before use, not only at preventive maintenance, at first use or after someone gets a shock.
- Remove damaged equipment from service. No one may use it until repairs and testing make it safe. Tag it and send it to Biomedical or Clinical Engineering.
- Electrical tape over a frayed cord is not a repair, and "it still works" is not the same as safe.
- Plugs may not be altered (for example, filed down), and adapters that break the grounding connection are prohibited. A pump missing its grounding prong is removed from service, not run through a two-prong adapter, even briefly.
Maintenance, function checks and tracking
- CMS requires hospital equipment to be inspected, tested and maintained for safety, with the maintenance documented, and the work overseen by qualified staff.
- If a preventive maintenance (PM) label is past due, hold the unit and send it to Clinical Engineering. A cleaning function check is not PM, and technicians do not write new due dates.
- CMS expects all hospital equipment to be listed in an inventory with its maintenance record, whether it is owned or leased. A rented specialty bed is added and held to the same standards.
- Before issue, check function according to the manufacturer's IFU and facility policy, not a coworker's routine or the steps for a different model.
- Barcoding pumps and scanning them at issue and return lets the department find specific units quickly, such as those named in a recall or due for maintenance. A scan records location; it does not prove the unit was cleaned or replace PM.
Sending equipment out for repair
- OSHA requires equipment that may be contaminated to be examined before servicing or shipping and decontaminated as needed.
- If a part cannot be decontaminated, attach a readily visible label stating which portions remain contaminated, and tell the people who will service it.
- Do not ship it as is and leave it to the vendor, do not take the device apart to discard the part, and do not hold it indefinitely in-house.
Key terms
- Noncritical item
- Equipment that touches only intact skin or no patient at all; needs cleaning plus low-level disinfection.
- Spaulding classification
- A system sorting devices as critical, semicritical or noncritical by infection risk, which sets the processing level.
- Low-level disinfection
- Killing most vegetative bacteria and some fungi and viruses; the level CDC recommends for noncritical items.
- Contact time
- How long a disinfectant must stay wet on a surface to work, as stated on its label.
- EPA-registered disinfectant
- A surface germicide whose label claims and directions are approved by the Environmental Protection Agency and legally binding.
- Cross-contamination
- Spread of microbes from soiled items or surfaces to clean ones.
- Single-patient-use item
- A part, such as a compression sleeve, meant for one patient and discarded rather than reprocessed.
- Biomedical (Clinical) Engineering
- The hospital department that inspects, maintains and repairs medical equipment.
- Preventive maintenance (PM)
- Scheduled inspection, testing and servicing of equipment by qualified staff, documented on a label or record.
- Grounding prong
- The third plug pin that keeps equipment connected to ground; altering it or bypassing it with an adapter is prohibited.
- Function check
- A test before issue that equipment powers on and works as its manufacturer's instructions describe.
- Equipment inventory
- A list of all hospital equipment, owned or leased, with each item's maintenance record.
Common mistakes
- Calling pumps or blood pressure cuffs semicritical, or thinking noncritical means exempt from processing. They are noncritical and still need cleaning and low-level disinfection.
- Spraying disinfectant over visible soil, or judging contact time by how fast the surface dries. Clean first, then keep it wet for the label's contact time.
- Soaking a pump or steam sterilizing powered equipment without checking the IFU for immersion and heat tolerance.
- Issuing equipment with a frayed cord, a missing ground pin or an overdue PM label, or fixing it with tape, an adapter or a hand-written date. Remove it from service and send it to Clinical Engineering.
- Skipping gloves because blood has dried, or thinking a barcode scan proves a unit was cleaned or replaces maintenance.
- Shipping contaminated equipment to a vendor as is. Decontaminate what you can and label the parts that remain contaminated.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated), PDF; CDC, Guidelines for Environmental Infection Control in Health-Care Facilities (2003, updated 2019), PDF; CDC, Core Infection Prevention and Control Practices for Safe Healthcare Delivery in All Settings; CDC, Disinfection and Sterilization Guideline: Sterilizing Practices; OSHA, 29 CFR 1910.1030 Bloodborne Pathogens; OSHA, 29 CFR 1910.334 Use of equipment (electrical); CMS, S&C 14-07-Hospital, Hospital Equipment Maintenance Requirements (2013); FDA, Benefits of a UDI System; Winnipeg Regional Health Authority, Transportation, Distribution, and Storage of Contaminated, Clean and Sterile Medical Devices (operational directive, 2023)