This topic covers why cleaning comes first, point-of-use care, manual cleaning at the sink, detergents and how to dilute them, ultrasonic cleaners, washer-disinfectors and cart washers, rinse water, and single-use and loaner devices. For most questions the deciding fact is the device or detergent manufacturer's instructions (IFU).
Why cleaning comes first
- Soil left on a device, organic or inorganic, acts as a shield between microbes and the disinfectant or sterilant.
- A dirty device cannot be reliably disinfected or sterilized; a longer sterilizer cycle does not make up for it.
- Start cleaning as soon as practical after use, because dried or baked-on soil is harder to remove and can make disinfection or sterilization fail.
- Every used item that arrives in decontamination is treated as contaminated, so PPE is worn to handle it.
- After cleaning, every instrument is still inspected and must be visibly clean.
Point of use and receiving
- Point-of-use care does not replace cleaning; it is the first step of cleaning.
- Keep soiled instruments open and moist with an enzymatic spray, a towel wet with water, or a foam or gel made for this purpose.
- Do not use saline to keep instruments moist or to soak them; saline (like dried blood) can damage stainless steel and other devices.
- A loaner set from a vendor enters the same cycle as the facility's own devices: clean, inspect, package and sterilize it per the IFU, whatever another site says it did.
Manual cleaning at the sink
- Take devices with removable parts apart as far as the IFU allows, and keep all the parts together so they can be reassembled.
- CDC names two essentials of manual cleaning: friction (scrubbing with a brush or cloth) and fluidics (fluid under pressure to flush channels).
- Fully submerge immersible items and scrub them below the water surface so aerosols do not form.
- Do not scrub under running water or above the water line; both spray contaminated droplets.
- Use soft nylon brushes; metal brushes and abrasives scratch instruments.
- A lumen brush should match the lumen's diameter so the bristles touch the walls, and be long enough to come out the far end.
- Reusable brushes must be cleaned and then sterilized or high-level disinfected between uses, unless disposable brushes are used.
Detergents, enzymes and dilution
- Neutral or near-neutral pH detergents are most gentle on device materials and still remove soil well.
- Enzymatic cleaners usually contain proteases that break down proteins such as blood and pus; some add lipases for fat and amylases for starch.
- Enzymatic cleaners are not disinfectants: they kill nothing, so items stay contaminated.
- Because they do not stop microbes growing, discard enzymatic solutions after each use, for example after each endoscope.
- Rinse enzymes off devices; residue can cause reactions such as fever.
- Per CDC, a neutral pH detergent with enzymes is the best choice for delicate instruments, especially flexible endoscopes; alkaline cleaners dissolve protein and fat well but can corrode.
- Household dish or laundry detergents are not suitable for medical devices.
- Dilution, water temperature and contact time come from the detergent label; some products are designed for a specific temperature.
- A stronger mix is not necessarily more effective, and water that is too hot sets protein: WHO says protein coagulates above 60°C (140°F).
Example
The label calls for a 1% solution. The sink holds 10 liters and the detergent comes full strength (100%). How much detergent goes in?
Solution
Detergent volume = concentration needed × sink volume in mL ÷ supplied concentration = 1 × 10,000 ÷ 100 = 100 mL. Add 100 mL and fill to 10 liters. Check: 10 mL would be 0.1% (too weak) and 1,000 mL would be 10% (too strong).
Ultrasonic cleaners
- Sound waves pass through the solution and form bubbles that grow, become unstable and implode; this cavitation pulls soil out of box locks, serrations, hinges and lumens.
- Remove gross soil first, open the instruments, submerge them fully and fill the lumens.
- Degas the unit every time it is filled: fill it, close the lid and run a cycle for 5 to 10 minutes, because dissolved gas weakens cavitation.
- Change the solution daily and whenever it is visibly dirty; used solution can hold bacteria and endotoxin.
- Close the lid before running a cycle, then rinse and dry the instruments afterward.
- Check each device's IFU first; some powered or lensed devices must not go in an ultrasonic cleaner, and fitting in the basket does not make them safe.
Washer-disinfectors and cart washers
- Washer-disinfectors clean by impingement: pressurized water sprays the soil off, helped by detergent and heat.
- WHO's typical cycle: pre-rinse, enzymatic wash, detergent wash, lubrication, then a final rinse hot enough to disinfect, ideally with deionized water to prevent spotting.
- Open hinged instruments fully, take items apart as far as possible, and never stack instruments or trays.
- Do not overload racks; place multi-level trays separately and remove all lids.
- Inspect the spray arms daily and keep them clear.
- A commercial test soil that mimics dried blood can check the washer; a failure points to a machine fault or a chemical not feeding in, but a pass does not prove instruments are clean.
- Check and document each cycle's parameters.
- Cart washers clean carts, rigid containers and basins; their cycle lacks the enzymatic wash and lubrication, so instruments go in only if the equipment maker has validated it.
Rinse water and TASS
- Rinse after cleaning to remove loosened soil and detergent, which could react with the disinfectant or sterilant.
- WHO: give lumens of intravascular or intrathecal devices a final rinse with sterile, pyrogen-free water or reverse osmosis water; distilled water may still carry microbes or pyrogens.
- Toxic anterior segment syndrome (TASS) is acute, noninfectious inflammation of the front of the eye, usually within 24 hours of surgery such as cataract removal.
- Most TASS cases trace to contaminants on instruments from poor cleaning, or to substances that enter the eye during surgery.
- In a 2006 Maine cluster, changes included rinsing with sterile distilled water instead of tap water after the ultrasonic bath, changing the bath twice daily instead of once, and stopping the enzymatic cleaner in the bath.
Single-use devices
- A single-use device is meant for one use, or one patient during one procedure, and then discarded.
- Single-patient use is a different label; single-use items are not cleaned and reused just because they look clean.
- FDA holds a hospital or third party that reprocesses single-use devices to the same rules as the original manufacturer: premarket clearance, registration and listing, adverse event reporting, tracking, recalls and labeling.
Key terms
- Point-of-use treatment
- Keeping used instruments open and moist right after the procedure; it starts but never replaces cleaning.
- Friction
- Rubbing or scrubbing soil off with a brush or cloth; one of two essentials of manual cleaning.
- Fluidics
- Fluid forced under pressure through channels to flush out soil, especially where a brush cannot pass.
- Enzymatic cleaner
- Detergent with proteases (and sometimes lipases or amylases) that digests soil but kills no microbes.
- Neutral pH detergent
- Cleaning solution that is neither acidic nor alkaline; with enzymes it is CDC's preferred choice for delicate instruments and flexible endoscopes.
- Alkaline cleaner
- High-pH detergent that dissolves protein and fat efficiently but can corrode devices.
- Cavitation
- Formation and implosion of tiny bubbles from sound waves, which pulls debris off instruments in an ultrasonic unit.
- Degassing
- Running a freshly filled ultrasonic unit with the lid closed for 5 to 10 minutes to remove dissolved gas.
- Impingement
- Pressurized water spray that physically knocks soil off; the main cleaning action of a washer-disinfector.
- Thermal rinse
- Final hot-water step of a washer-disinfector cycle that disinfects by heat.
- Test soil
- Commercial product mimicking dried blood, used to check that a washer-disinfector is cleaning properly.
- Cart washer
- Machine for carts, rigid containers and basins; lacks enzymatic wash and lubrication steps.
- Pyrogen-free water
- Water free of fever-causing substances such as endotoxin; required for final rinsing of intravascular device lumens.
- TASS
- Toxic anterior segment syndrome: acute, noninfectious eye inflammation after anterior segment surgery, often from instrument residue.
- Single-use device
- Item intended for one use or one patient in one procedure, then discarded.
- Instructions for use (IFU)
- Manufacturer's written directions for reprocessing a device or using a chemical; they set disassembly, dilution, temperature and method.
Common mistakes
- Point-of-use wiping or spraying does not disinfect and does not replace cleaning, and saline is never used to keep instruments moist.
- Do not scrub under running water or above the water line; scrub immersible items below the surface. Lumen brushes must match the lumen diameter, not be the smallest or a forced oversize.
- Enzymatic cleaners are not disinfectants, must be measured and rinsed off, and are discarded after each use. More detergent, hotter water or a stronger mix is not better; follow the label.
- Cavitation is the ultrasonic cleaner; impingement is the washer-disinfector. Degas after every fill and change cloudy solution now, not at the end of the shift.
- In washers, keep hinged instruments open, do not stack or overload, and remove lids. A failed test soil means a washer or chemical-feed problem, not a sterility result, and a cart washer is not used for instruments unless validated.
- Distilled water may still contain microbes or pyrogens. A loaner marked 'processed' elsewhere still gets full reprocessing, and a hospital that reprocesses single-use devices is treated by FDA as a manufacturer.
Checked against: CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated 2024), print version; WHO/PAHO, Decontamination and Reprocessing of Medical Devices for Health-care Facilities (2016); CDC MMWR, Toxic Anterior Segment Syndrome After Cataract Surgery, Maine, 2006; FDA, Reprocessing of Single-Use Devices (archived page)