This topic covers which agency or organization controls what, why the manufacturer's instructions govern reprocessing, what sterilization records must hold, how device problems are reported to FDA, and the main quality improvement methods. Match each body to its job, and know the deadlines and record contents CDC and FDA give.
Who regulates what
- **FDA** (its Center for Devices and Radiological Health) regulates firms that make, repackage, relabel or import medical devices sold in the U.S. Devices reach the market through 510(k) clearance or premarket approval (PMA).
- FDA also oversees the liquid sterilants and high-level disinfectants sold for use on critical and semicritical devices.
- **EPA** registers disinfectants used on environmental surfaces such as counters and floors, under the federal pesticide law (FIFRA).
- The split dates from a 1993 FDA-EPA agreement and the 1996 Food Quality Protection Act, which moved device sterilants and high-level disinfectants out of EPA's hands.
- **OSHA** exists to ensure safe and healthful working conditions, through rules such as the Bloodborne Pathogens and Hazard Communication standards.
- **CDC** writes guidelines; it does not regulate or clear devices.
- **CMS** sets the Conditions of Participation, the health and safety rules an organization must meet to begin and keep taking part in Medicare and Medicaid. Hospital conditions are in 42 CFR Part 482.
- **Accrediting organizations** such as The Joint Commission are approved by CMS. A facility they accredit can show it meets Medicare rules through **deemed status**, granted after CMS reviews the organization's standards and survey process.
Standards, labels and the IFU
- Consensus standards, such as those from AAMI, are written by standards development organizations. FDA calls them voluntary unless a standard is written into regulation.
- AAMI is a standards development organization, not a government agency, so its standards are not federal rules.
- Every EPA-registered disinfectant label states that using the product in a way that does not match its labeling breaks federal law. Use the exact dilution and contact time on the label; both are binding, for every user.
- FDA explains that a reusable device can be safely used again when its reprocessing instructions are followed completely and correctly after every use.
- So the manufacturer's written instructions for use (IFU) govern how a device is cleaned and processed. Not a similar brand, not a coworker's habit and not the department's usual setting. If the IFU is unclear, contact the manufacturer.
- When two IFUs conflict, such as an endoscope's and its automated reprocessor's, CDC says to compare both and resolve the conflict before use. Neither one wins automatically.
Sterilization records and traceability
- For each cycle CDC lists: sterilizer type and cycle, load identification number, load contents, exposure parameters (time and temperature), operator name or initials, and the mechanical, chemical and biological monitoring results.
- Date and initials alone are not enough, and biological indicator results are only one part of the record.
- The cycle record describes the load, not patients; CDC adds patient identifiers only when an implant has to be flash sterilized.
- Keep mechanical, chemical and biological records as long as standards (for example, 3 years), state and federal rules and statutes of limitations require. Not just until the indicator is read or the next service visit.
- Each package carries a load number so its items can be traced to their cycle and recalled if a problem with that load is found. It is not for billing, shelving order or skipping inspection.
Reporting device problems to FDA
- **MedWatch** is FDA's program for safety information and adverse event reports. Health professionals can send voluntary reports on Form FDA 3500.
- User facilities (such as hospitals) must report a suspected device-related **death** to both FDA and the manufacturer within 10 work days of learning of it.
- A device-related **serious injury** goes to the manufacturer within 10 work days, or to FDA if the manufacturer is unknown.
- Facilities also send FDA a yearly summary of these reports. It does not replace the individual reports.
- If a technician suspects a reprocessed device (for example, retained debris) harmed a patient, report it through the facility's internal procedure. Include make, model, catalog and serial number and a full description of what happened.
- Do not quietly reclean and reissue the device, and do not throw it away; it is needed for evaluation. Telling the patient is handled under facility policy, not by the technician alone.
Staff competency and using data
- CDC: give hands-on training, supervise each reprocessing task until competency is documented, and test competency at the start of employment and regularly after that, such as once a year.
- Competency checks are the employer's program; they are separate from personal certification and do not wait for a failure.
- CDC says flash (immediate-use) steam sterilization should not be used for convenience, to save time, or in place of buying more instrument sets.
- If an audit shows the same set being flashed nearly every day because only one exists, the fix is more inventory, not a new cycle or removing the sterilizer.
Quality improvement methods
- **PDSA (Plan-Do-Study-Act):** Plan the change and its goal; Do a small-scale test; Study the results; Act by adopting the change if it worked. Good for quick, small gains.
- **Lean:** removes steps, processes or inputs that add no value; suits workflows with bottlenecks and rework. Mapping a case cart process and cutting wasted steps is Lean.
- **Six Sigma:** uses data to reduce variation and defects through DMAIC (Define, Measure, Analyze, Improve, Control); suits repetitive, high-volume work such as tray assembly.
- **Root cause analysis (RCA):** a structured look back at a serious event to find the system weaknesses that made the error likely, rather than blaming one person. The Joint Commission requires RCA for sentinel events.
Key terms
- FDA
- Federal agency that regulates medical device makers and clears sterilants and high-level disinfectants for devices.
- EPA
- Federal agency that registers disinfectants for environmental surfaces such as floors and countertops.
- OSHA
- Federal agency that sets and enforces rules to keep workers safe and healthy on the job.
- Conditions of Participation
- CMS health and safety requirements an organization must meet to take part in Medicare and Medicaid.
- Deemed status
- Recognition that a facility accredited by a CMS-approved organization meets Medicare requirements.
- Consensus standard
- A voluntary practice document written by a standards development organization such as AAMI.
- Instructions for use (IFU)
- The manufacturer's validated written directions for using, cleaning and processing a specific device.
- FIFRA
- The federal pesticide law under which EPA registers surface disinfectants and makes their labels legally binding.
- Load number
- A code on each package linking it to the sterilizer cycle that processed it, for tracing and recalls.
- MedWatch
- FDA's program for safety information and adverse event reports; voluntary reports use Form FDA 3500.
- User facility reporting
- FDA rule requiring hospitals to report device-related deaths and serious injuries within 10 work days.
- Competency testing
- Documented checks that a worker can do each task correctly, done at hire and regularly after.
- PDSA
- Plan-Do-Study-Act: an improvement cycle built on small tests of change before wider adoption.
- Lean
- An improvement method that raises efficiency by removing steps that add no value.
- Six Sigma
- A data-driven improvement method that reduces variation and defects using Define-Measure-Analyze-Improve-Control.
- Root cause analysis (RCA)
- A structured review of a serious event to find underlying system problems instead of individual blame.
Common mistakes
- Mixing up agencies: EPA registers surface disinfectants, FDA clears high-level disinfectants and sterilants for devices, OSHA protects workers, CMS runs Medicare rules, and CDC only writes guidelines.
- Treating a disinfectant label or IFU as a suggestion. The label's dilution and contact time are legally binding, and the device IFU, not a similar device or a coworker, governs reprocessing.
- Thinking AAMI standards are federal regulations enforced by inspection. FDA describes consensus standards as voluntary unless written into regulation.
- Getting the reporting rules wrong: a device-related death goes to both FDA and the manufacturer within 10 work days, and the annual report does not replace it.
- Throwing away or quietly recleaning a device suspected of harming a patient. Report it internally with device details and keep it for evaluation.
- Confusing the improvement methods: Do is the small pilot test in PDSA, Lean removes waste, Six Sigma uses DMAIC to cut variation, and RCA looks for system causes, not someone to blame.
Checked against: FDA, Overview of Device Regulation; OSHA, About OSHA; OSHA, 29 CFR 1910.1200 Hazard Communication; OSHA, 29 CFR 1910.1030 Bloodborne Pathogens; CMS, Conditions for Coverage & Conditions of Participation; CMS, Accrediting Organizations (AOs); CMS, S&C 14-07-Hospital, Hospital Equipment Maintenance Requirements (2013); FDA, Division of Standards and Conformity Assessment; FDA, Reprocessing of Reusable Medical Devices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated), PDF; FDA, Medical Device Reporting (MDR): How to Report Medical Device Problems; FDA, Mandatory Reporting Requirements: Manufacturers, Importers and Device User Facilities; FDA, Reporting Problems with Reusable Medical Devices or Reprocessing; Atia MA et al., Quality monitoring in colonoscopy: Time to act. World J Gastrointest Endosc 2015 (PMC4400621); AHRQ PSNet, Root Cause Analysis (primer)