This topic covers where and how sterile items are stored, how long they stay usable under event-related and time-related shelf life, what to do with a torn, wet, dropped or expired pack, and how sterile items travel to the units. Most questions turn on one idea: anything that can damage or wet the package can end its sterility.
The storage room
- CDC describes sterile storage as a limited-access area. Only authorized staff who need to work there should go in.
- Fewer people means less handling, less dust and fewer events that can damage packs. A sign-in log for visitors or vendors does not make general traffic acceptable.
- CDC 2008: room temperature is controlled and can reach up to 75°F, and relative humidity should stay at or below 70%.
- Closed or covered cabinets are ideal, but CDC says open shelving may be used. Neither is banned.
- Never store sterile supplies under sinks or anywhere they could get wet. Packaging type and closed cabinet doors do not protect against leaks.
Clearances on the shelves (CDC 2008)
- Floor: store supplies 8 to 10 inches above the floor.
- Ceiling: at least 5 inches below the ceiling in general.
- Sprinkler heads: at least 18 inches below a sprinkler head.
- Outside walls: 2 inches away from outside walls.
- Why: the gaps let air circulate, make the area easy to clean and keep the room within local fire codes (sprinklers must be able to spray freely). They are not about reach or keeping packs warm.
- Department policy may give its own figures. When a question names CDC 2008, use the numbers above.
Keeping packs intact on the shelf
- Corrugated cardboard shipping boxes stay outside sterile storage. They are not stored on any shelf there, open or closed, and are not used as spacers between items.
- Do not stack heavy trays on top of peel pouches or light packs. CDC says to store items so packaging is not punctured or bent, and rigid containers are heavy too.
- Handle packs as little as possible. CDC notes the chance of contamination goes up as handling increases, and this applies to wrapped trays as well as pouches.
- Rotate stock first in, first out: put new stock behind the older stock of the same item so the older items are used first. Do not stack new stock on top or keep it on a separate shelf.
- Rotation helps use items in date order. It does not repair a damaged or wet pack.
Event-related and time-related shelf life
A facility uses one of two approaches, and each gives a different answer to "is this pack still good?"
- **Event-related:** a properly packaged sterile item stays sterile until something happens to it, such as a tear, wetting, a broken seal, a puncture or crushing. CDC cites data showing no trend toward contamination over time in covered storage.
- Event-related does not mean "never expires no matter what". Room conditions and storage practices still matter, because they decide whether events happen.
- Event-related also does not mean sterility slowly fades each week. It is tied to events, not to the calendar.
- **Time-related:** each pack is labeled with an expiration date. Once that date passes, the pack is not issued; it is reprocessed, even if the wrap looks perfect.
- Never write a new date on an expired pack, and never send it to a unit just because the unit will open it soon.
Inspect before use: torn, wet or dropped packs
- Check every package before issue. If it is torn, wet or punctured, pull it from stock and repackage and reprocess it before use.
- Indicator tape over a tear does not restore the barrier. Moving a damaged tray to the front of the shelf does not fix it either.
- Check a dropped package for damage to its packaging, and to the contents if they are breakable.
- If the dropped item sits in a heat-sealed impervious plastic pouch whose seal is still whole and undamaged, CDC says it is not considered contaminated and need not be reprocessed.
- So a fall alone is not an automatic discard or reprocess; inspection decides. Never just wipe a dropped pack and send it out unchecked.
Transporting and delivering sterile items
- Perform hand hygiene before handling or restocking clean or sterile supplies.
- Use a cart with a solid base that is covered or closed while it is outside the department. An open wire cart leaves packs exposed to dust, splashes and contact.
- Clean or sterile items and contaminated items never share a cart, and one person should not handle both at the same time. If asked to bring back used instruments, get a separate cart.
- A towel over a soiled bin, or carrying the bin by hand while pushing the sterile cart, is still mixing clean and dirty.
- Off-site services: a vehicle compartment that carried contaminated devices must be cleaned and disinfected before it carries clean or sterile items, and sterile items travel covered or contained. Airing out or lining the floor with linen is not enough.
Key terms
- Sterile storage
- A limited-access area with controlled temperature and humidity where processed packs are kept until issue.
- Event-related shelf life
- The practice of treating a pack as sterile until damage, wetting or another event compromises it.
- Time-related shelf life
- The practice of labeling packs with an expiration date and reprocessing them once that date passes.
- Package integrity
- The condition of a wrap, pouch or container being whole, dry and sealed, with no tears or punctures.
- Impervious packaging
- Material, such as heat-sealed plastic, that does not let moisture or microbes pass through.
- Stock rotation
- Placing newer items behind older ones so the oldest are used first; also called first in, first out.
- First in, first out (FIFO)
- Using the oldest stock of an item before newer deliveries of the same item.
- Open shelving
- Uncovered racks for supplies; allowed by CDC, though closed or covered cabinets are ideal.
- Sprinkler clearance
- The 18-inch space CDC calls for between stored items and a ceiling fire sprinkler head.
- Corrugated cardboard
- Ridged material used for shipping boxes; such boxes are never taken into sterile storage areas.
- Closed cart
- A transport cart with walls, doors or a cover that protects its load while outside the department.
- Reprocessing
- Repeating cleaning, inspection, packaging and sterilization so an item can be used safely again.
Common mistakes
- Mixing up the CDC clearances: 8 to 10 inches from the floor, 5 inches from the ceiling, 18 inches from a sprinkler head, 2 inches from outside walls.
- Thinking event-related shelf life means a pack never needs checking or can be issued with a tear. A tear, wetting or broken seal is exactly the event that ends sterility.
- Issuing, redating or sending to the OR a pack past its date under time-related shelf life. Once the date passes it is reprocessed, whatever the wrap looks like.
- Discarding or automatically reprocessing a dropped heat-sealed impervious pouch. Inspect it; if the seal and contents are intact, it can stay in use.
- Believing a closed cabinet door, plastic pouches or FIFO rotation make storage under a sink acceptable, or that cardboard boxes are fine if kept closed.
- Putting soiled returns on the sterile delivery cart, even on a lower shelf or under a towel. Use a separate cart.
Checked against: CDC, Disinfection and Sterilization Guideline: Sterilizing Practices; CDC, Guideline for Disinfection and Sterilization in Healthcare Facilities (2008, updated), PDF; Winnipeg Regional Health Authority, Transportation, Distribution, and Storage of Contaminated, Clean and Sterile Medical Devices (operational directive, 2023); Wikipedia, Stock rotation